the appliance is properly grounded. Do not modify the plug provided with the appliance – If it will not
fit the outlet, have a proper outlet installed by an electrician
This equipment has a grounding plug that looks like the plug illustrated in sketch A below. A temporary
adaptor, which looks like the adaptor illustrated in sketches B and C, may be used to connect this plug
to a 2 pole receptacle as shown in sketch B if a properly grounded outlet is not available. The
temporary adaptor should be used only until a properly grounded outlet can be installed by a qualified
electrician. The green color rigid ear, lug and the like, extending from the adaptor must be connected
to a permanent ground such as a properly grounded outlet box cover. Whenever the adaptor is used, it
must be held in place by a metal screw.
Turning on the power switch will enter into below user interface:
6. Parameters setting
• User is required to set the room size, the residual dose volume of the disinfectant, then click "Start",
Start Delay countdown will start and when it ends the equipment will start operation, after the
sterilization time duration is over, the equipment stops operation
• The formula of Sterilization Time is: Room Volume× 5ml ÷ 16ml
• 5ml: Every 1m³ space requires 5ml of disinfectant
• 16ml: equipment is able to spray 16ml of disinfectant per minute
• The parts that can be clicked on the touch screen are Room Volume, Residual Dose, Countdown
Function logic: