The bulb and bulb-protector can now be installed. Point the lamp face
upward and drop in the bulb-protector
. Next, screw the bulb in.
The bulb
should fit snugly against the inside rim of the bulb-protector
.
4.
Replace the filter assembly and front bezel.
You
are now ready to oper-
ate the lamp.
5.
Plug the lamp into an appropriate electrical outlet and switch the lamp
on. Allow
approximately five minutes to reach full intensity .
Notes:
When use is intermittent, it is preferable to leave the lamp burning.
This assures longer bulb life and avoids delay in restarting.
Restarting does
not take place immediately as the mercury vapor within the bulb must cool.
Delay is typically 4 to 8 minutes.
Any interruption in the power supply
, such
as an excessive drop in voltage, will extinguish the arc. Restarting typically
will be delayed while the mercury vapor cools. Full voltage must be restored
to permit the arc to strike.
Replacement Parts List
B-100SP Ultraviolet Lamp (1
15 VOL
T/60Hz)
PART
NO
.
PART
NAME
QTY
14-0070-02
END CAP
1
23-0051-02
FRONT BEZEL
1
38-0168-01
LENS BEZEL
ASSEMBL
Y
1
34-0049-01
BULB, SELF-BALLASTED
1
17-0057-01
BULB PROTECTOR
1
77-0007-02
LAMP FIXTURE
ASSEMBLY
1
Accessories
PART
NO
.
PART
NAME
34-0022-01
BULB, FLOOD
89-0068-01
FINGER GUARD
19-0124-01
LAMP VISOR
CAUTION
1.
Wear hand and eye protection if filter glass is cracked. Replace broken
filter as soon as possible.
2.
To avoid electrical shock hazard, unplug lamp before servicing.
3.
Bulb runs very hot; allow unit to fully cool before servicing.
4.
Bulb cont
ains mercury
. If bulb breaks, wear gloves to remove.
If rep
air of the lamp is necessary
, remove bulb prior to shipping to the repair
facility. A
Return Authorization Goods (RGA)
number must be obt
ained
from UVP to return the item.
23-0051-02
38-0168-01
14-0070-02
77-0007-02
34-0049-01
17-0057-01