N S T A L L A T I O N
Initial Setup
Unpack the Finesse
®
unit from its packing material. Retain the pack-
ing material and box for future use.
Inspect the unit for any visible damage or missing accessories. If
damage is found, contact your Utah Medical Products' representative
for replacement parts.
The unit should be placed on a flat, level surface at working height
within six feet of the operating area.
Connect the footswitch (if used).
Plug the Finesse
®
unit into an appropriately rated wall outlet.
Functional Checkout
1.
Install handswitch pencil into the three-point connector on
front panel. If a non-switching "footswitch" pencil is used,
install this pencil into the port marked with the "active port"
icon on the front panel.
2.
Toggle the main power switch to the "on" position. Three
things should happen:
•
the green lamp in the power switch should light,
•
the red alarm lamp on the front panel should light, and
•
an interrupted alarm tone should sound.
3.
If the three indicators listed in step 2 above do not occur,
check to verify that:
•
the power cord is securely plugged in on both ends,
•
the three fuses on the rear panel of the unit are installed,
•
these fuses are not blown, and the room outlet is active.
4.
Connect a dispersive pad with a single contact surface to the
dispersive pad receptacle on the front panel. The red alarm
lamp should go out and beeping should cease.
5.
Briefly operate the cut control on the handswitch. The yellow
"cut" lamp should light, a tone should be emitted from the
internal speaker, and the vacuum pump should come on and
stay on approximately five seconds after the cut control button
is released. If none of this occurs try another handswitch. If
the vacuum motor does not come on, check the position of
the vacuum control switch and the 3-amp fuse on the back
panel.
6.
Repeat step 5 above with the coag switch rather than the cut
switch. All operation should be the same except that the blue
"coag" lamp illuminates rather than the yellow "cut" lamp.
7.
Repeat steps 5 and 6 above using the cut and coag pedals of
the foot switch. All indications should be the same as they
were using the hand switch.
If any of the above items do not check good, please contact Utah
Medical Products Customer Service for assistance.
Maintenance
Daily Maintenance
At the end of any day that the Finesse
®
unit is used, the primary fil-
ter, catalog number ESU-501, should be changed. Using a gloved
hand, remove the filter from the front panel by gently pulling and
twisting the filter housing. Discard the filter pack with other medical
disposables.
Cleaning
The Finesse
®
system exterior may be wiped clean with alcohol or
household spray cleaner applied to a cloth. Do not use acetone.
Do not apply cleaning agent directly to the unit.
The loop and ball electrodes, pens, dispersive pads, and speculum
Finesse Service Manual
1
Содержание Finesse ESU-110
Страница 4: ......
Страница 10: ...Finesse Service Manual 6 ...
Страница 27: ...Mains Power Wiring Diagram Finesse Service Manual 23 ...
Страница 28: ...Assembly Diagram Finesse Internal View Finesse Service Manual 24 ...
Страница 29: ...RF Power Board Assembly Drawing Finesse Service Manual 25 ...
Страница 30: ...Logic Board Assembly Drawing Finesse Service Manual 26 ...
Страница 31: ...Display Board Assembly Drawing Power Supply Board Assembly Drawing Finesse Service Manual 27 ...
Страница 32: ...System Functional Diagram Finesse Service Manual 28 ...