Entering the Abdomen: 

The ‘Half Closed’ Position

 

A feature of the Sterishot II Applicator is a white visual guide located within 

the trigger housing, designed to help signal the ‘Half Closed’ position and 

hence avoid prematurely closing or distorting the Filshie Clip so that it may 

fail to close properly. To ensure the ‘Half Closed’ position is not exceeded, the 

edge of the black polished section of the trigger should not pass beyond the 

white visual guide located inside the black trigger housing. 

Application of the Filshie Clip 

 

•  Identify and inspect the Fallopian tube thoroughly.

•  Ensure that the Filshie Clip can accommodate the whole diameter of the 

Fallopian tube.

•  Locate the Filshie Clip over the isthmic portion of the Fallopian tube, 1-2cm 

from the cornu.

•  Having established the best location for the Filshie Clip after insertion 

through the trocar, the applicator should be re-opened and advanced a few 

millimeters to move the Fallopian tube gently to the back of the Filshie Clip, 

close to the hinge.

•  Close the Filshie Clip into position by applying firm, but gentle pressure on 

the trigger in a smooth action until the trigger reaches its mechanical stop.

•  When the Filshie Clip is secured in position, gently release the trigger and the 

Filshie Clip will automatically free itself from the applicator ‘Clip Tray’.

•  Do not use an abrupt closure action or the tube might be transected. Should 

this occur, apply a second Filshie Clip on the proximal (uterine) side of the 

transection.

Digital photography and video recording of the closure process are encouraged to 

support the patient record case file.

Removal of the Applicator from the Abdomen 

 

Before the Sterishot II Applicator can be withdrawn from the 

abdomen it must be returned to the ‘Half Closed’ position so it 

can pass through the cannula.

Correct Closure

Polished Trigger 

Section

White Visual Guide

Trigger Housing

Clip with applicator 

half closed

Clip open over tube

Clip closed on tube

Applicator jaws in

‘Half Closed’ position

(clip not prematurely closed or distorted)

Содержание Filshie Sterishot II

Страница 1: ...e Device for Permanent Female Sterilization Single Patient Use Filshie Clip Applicator IMPORTANT This guide is not intended to be a complete instruction Consult the Instructions for Use for complete indications contraindications warnings and precautions ...

Страница 2: ...ration of the applicator prior to use Users must be able to easily and fully compress the Sterishot II trigger to its stop Ensure the Clip is correctly loaded in the applicator 1 Check that the bottom portion of the Clip lies perfectly flat on the floor of the applicator Clip Tray not rising up at the back or the front 2 Check that the front end of the Clip lies directly behind the Clip Stop of th...

Страница 3: ...imeters to move the Fallopian tube gently to the back of the Filshie Clip close to the hinge Close the Filshie Clip into position by applying firm but gentle pressure on the trigger in a smooth action until the trigger reaches its mechanical stop When the Filshie Clip is secured in position gently release the trigger and the Filshie Clip will automatically free itself from the applicator Clip Tray...

Страница 4: ...ss Consistent highly effective results with a precalibrated Filshie Clip closure mechanism for each patient No post surgical cleaning sterile processing tracking between procedures or service required Order Reference Sterishot II Standard Laparoscopic Kit AVM 951 Sterishot II Minilaparotomy Kit AVM 951M Sterishot II Elite Kit w Bladeless Port AVM 954 Filshie Clips 10 pairs AVM 851J 7043 South 300 ...

Отзывы: