![Ushio MIDORI ULB-50sc Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/ushio/midori-ulb-50sc/midori-ulb-50sc_operating-manual_1064219005.webp)
Operating Manual
5
www.ushio.com
System Description and Operation
1. The light source power switch should be in the OFF position. Plug the
external 12vdc power supply into the 12vdc connector.
Figure 1.
2. Plug the external power supply cord into AC receptacle main power.
3. Plug the fiber-optic bundle into the light port and connect the opposite end
to the equipment being used.
Figure 2.
4. Turn the power switch to the ON position.
Figure 3.
5. Follow the instructions under
Touchscreen Operation
on page 7 to set the
desired UV output level.
6. Turn unit OFF when not in use.
Figure 3.
7. This unit is equipped with a fiber sense interlock feature. The device will not
emit light unless fiber cable is completely inserted into the front adapter.
Please ensure fiber cable is fully seated into light port.
8. To ensure consistent UV output use a 3/32” hex wrench and tighten the set
screw shown in
Figure 4
onto the light guide.
Power Switch
Set Screw
Figure 1. Back View
Figure 2. Fiber-Optic Bundle
Figure 3. Front View
12vdc
Power Plug
Light Port
Caution: Hot Surface - Light guide
ferrule may be hot when removed.
Do not touch. Allow to cool.
Figure 4. Bottom View