URCERI IR-817 Скачать руководство пользователя страница 2

Calibration Temperature at 300

Set the standard temperature source at 300

, aim IR sensor target at the standard temperature source

center, press the trigger button and move the sensor to find the hottest temperature site, and then press the
MODE button to save the adjusted data. Press the MODE button again to enter into Type K Temp.
Calibration mode.

Type K Temp. Calibration Mode

Input 0mV

from the calibration device, then press the trigger button

one time

, and then press MODE

button to save the data.

Input 4.095mV

the second time, then press the trigger button

one time

, and then

press MODE button to save the data.

After finishing the calibration, remove the battery to exit the calibration mode.