Calibration Temperature at 300
℃
Set the standard temperature source at 300
℃
, aim IR sensor target at the standard temperature source
center, press the trigger button and move the sensor to find the hottest temperature site, and then press the
MODE button to save the adjusted data. Press the MODE button again to enter into Type K Temp.
Calibration mode.
Type K Temp. Calibration Mode
Input 0mV
from the calibration device, then press the trigger button
one time
, and then press MODE
button to save the data.
Input 4.095mV
the second time, then press the trigger button
one time
, and then
press MODE button to save the data.
After finishing the calibration, remove the battery to exit the calibration mode.