background image

Page 9 

18. Exclusion 

 
If the equipment is used in a manner not specified by the manufacturer, the protection 
provided by the equipment might be impaired. 
This instrument is NOT suitable for use in explosion hazard environments. 
 
This equipment is not suitable and should not be used for medical applications. 
 

19. FAQ 

 

1. Q. Why does the pump stalls or runs slower than expected when new tubing is 

installed? 

A. A new tube is stiff when new so has to be shaped for some time before it gains it’s 
elasticity.  Run the pump in continuous mode for a total of approx 5 minutes with the 
clamp lever half open to start with, gradually moving the clamp lever to fully closed 
position. 

 

2. Q. I have entered a volume but nothing happens after pressing the GO button 

A. The volume you entered is too small for the tube size you use. The motor can not 
turn accurately a very small amount and the dispense is cancelled. Use a smaller tube 
diameter for low volume dispenses 

 

3. Q. What is the life of the tube? 

A. The life of a peristaltic tube depends on many variants such as speed, the reagent 
it is pumping and temperature. In general you should expect approximately 800 hours 
operation from one Pharmed tube and approx 150 hours from a Silicone tube. 

 

4. Q. Why does the displayed value  sometime ‘jump’ in double figures? 

 

A. This happens when dispensing small volumes and instrument in course calibration. 
Following the calibration procedure the instrument’s mathematical calculation divides 
the volume to be dispensed by the number of pulses received from the motor’s 
encoder and stores this value.  If the volume per rotor rotation is relatively large, the 
ratio of pulses to volume is coarse. In these circumstances the required volume input 
does not divide into stored values, and therefore the calculator will round up or down 
to the nearest whole figure.  

 

5. Q. Can I leave the unit switched ‘ON’ throughout the day? 

 

A. Yes, you can. The dispenser requires very little energy when the pump is not 
running so continuous power into the unit will not harm it. 

 

6. Q. Can I dispense hot reagents using the dispenser? 

 

A. Yes, the dispenser to works safely with media temperatures of up to 100°C. Be 
aware that the internal electronics could be damaged by any ingress of fluids or steam 
vapours. 

Page 8 

16. Remote control operation 

 
The instrument is fitted with a stereo type Ø2.5mm socket for remote control. 
Through this socket you can plug in the foot pedal which will provide the unit with a GO 
signal. 
 
The unit can be controlled from a computer via a USB relay 
controller which will convert an signal from your computer to a closed 
circuit on the remote wire connection. A USB relay controller is 
available from many electronic components suppliers. 
 
Full control of the unit is then possible using the software supplied 
with such relay. 
 
The stereo cable to the unit can be fitted with 3 core wire allowing the equipment to send a 
“busy” signal back to the controller. 

 

9100/9200 

Work State 

START 

3V-24V DC 

PC 

Busy dispensing 

Com (0V) 

17. Care & maintenance 
 

The control unit is maintenance free. The peristaltic tubes however require replacement as 
soon as excessive wear or a large variation in dispense volumes are noticed. 
 
The operational life of the tubes is a function of the speed, load and reagent being dispensed. 
 
Avoid running the tubes dry for longer than a few minutes. 
 
Check the peristaltic tubes weekly for signs of excessive wear and replace if required. 
 
Pump tubes which remain clamped in the pump will deform with time. Therefore, rotate the 
tube clip at the top of the pump to ‘open’ position in order to relieve the pressure whenever 
the pump is not being used for long periods or overnight. 

Содержание BOXER 9100

Страница 1: ...1 0mm complete with connectors Cat No 9000 504 Silicone Tubing IDØ2 0mm complete with connectors Cat No 9000 505 Silicone Tubing IDØ3 0mm complete with connectors Cat No 9000 510 Santoprene Tubing IDØ1 0mm complete with connectors Cat No 9000 501 Santoprene Tubing IDØ2 0mm complete with connectors Cat No 9000 502 Santoprene Tubing IDØ3 0mm complete with connectors Cat No 9000 518 Santoprene tubing...

Страница 2: ...ical compatibility chart 1 A Unpacking Remove the packing materials unpack the pump controller and the power supply unit Make sure that you have got all the components as specified in Box Contains last page of this manual Please contact your supplier if you notice any one of the components is missing or damaged Note Do not attempt to assemble a unit using damaged components Retain the packaging so...

Страница 3: ... pressure by lifting the pressure lever 3 C Replacing the tube Tube need replacing when worn out or wherever you intend to change the dispense volume range in which you work for smaller or larger dispense volumes Lift the pressure lever and replace the tube making sure that you have installed the tube centrally and that the tube is fully engaged by the tube guides Page 2 Pressure lever Clamp 4 The...

Страница 4: ...t be entered in R2 R3 Working with low volume dispenses with large diameter tube will reduce both precision and accuracy of dispenses Using the U D keys scroll to the screen displaying Range Enter Press Enter Using the U D keys choose the range corresponding to your tube diameter and press Enter The display will show the last volume you worked with at that particular range 8 Gravimetric pump calib...

Страница 5: ...ered volume in the reversed direction Page 7 13 Interval Dispense You can set the unit to dispense any volume for a number of times in predefined intervals This feature is useful in repeatable operations either in manual mode or when the dispense head is held by a robotic arm Intervals can be set between 0 5 sec to 100 minutes To set repeat dispensing 1 Using the U D keys scroll to the screen sayi...

Страница 6: ...de into stored values and therefore the calculator will round up or down to the nearest whole figure 5 Q Can I leave the unit switched ON throughout the day A Yes you can The dispenser requires very little energy when the pump is not running so continuous power into the unit will not harm it 6 Q Can I dispense hot reagents using the dispenser A Yes the dispenser to works safely with media temperat...

Страница 7: ...1 Santoprene tubing ID Ø2 00mm1meter running length Cat No 9000 506 Santoprene tubing ID Ø3 00mm1meter running length Cat No 9000 515 Silicone tubing ID Ø1 00mm1meter running length Cat No 9000 507 Silicone tubing ID Ø2 00mm1meter running length Cat No 9000 509 Silicone tubing ID Ø3 00mm1meter running length Cat No 9000 508 Tube clamp for continuous Ø1 0mm tube set of 2 Cat No 9000 601 Tube clamp ...

Отзывы: