How to Correct Common Failures F-94 and F-49

Why do these errors occur?

These two error codes are interrelated in the sense that the method for correcting them both is 
identical. These errors are typically caused by a total power loss to the pump that has forced 
the programing to revert back to the manufacturer’s default settings and will require a re-
programing in order to resume normal operation. The key part of the programing that needs to 
be reset is the time and date. The following process describes how to accomplish this task 
quickly and easily. 

Resetting the time and date to correct F-94 & F-49:

1. Plug the unit in and ensure that it is powered off
2. On the back side of the pump locate the “Panel Lock Out” button. It will be a small black

push button.

3. Press and hold the “Panel Lock Out” button then turn to the front side of the pump
4. On the front keypad, locate the “Stop” and “On/Off” buttons.
5. While continuing to hold the “Panel lock out button” also press the “Stop” button.
6. With these two buttons pressed and held, tap the “On/Off” button and continue holding

the two previously mentioned buttons.

7. Monitor the top screen of the pump. You are looking for it to say “Modify Config”. Once it

says this you can release the two buttons.

8. Now that you are in the “Modify Config” mode, locate the 3

rd

key down from the top of

the keypad that says “Tot Vol/ Status” and has the text “Next” lit above it.

9. This key will be used to cycle through the screens of the “Modify Config” mode.
10.Press this key approx. 18 times until you see the top screen say “Time” with some numbers

below it.

11.Don’t worry about what time the pump is currently saying we are now going to program

in the current time at your location in 24 hour or Military format. I.E. if it’s 2:34 pm where
you are then the time you’ll input is 1434.

12.Utilize the number keys to input the current time in the format listed, you’ll see that time

show up in the bottom screen.

13.Once the time is inputted correctly in the bottom screen, press the “Pri Start” key and you

should see the time move from the bottom screen to the top screen.

14.This indicates that your setting has been accepted. If you don’t see this move then you’ll

need to hit the clear button and repeat this process.

15.If you enter a wrong number at this step or the next simply use the “Tot Vol/Status” or

“Next” key to cycle through the menus back around to the option you are trying to
change.

16.After the time is properly inputted, press the “Tot Vol/Status” or “Next” key once more and

the screen should display “Date” followed by numbers for the date.

17.Don’t worry about what date the pump is currently saying. We are now going to program

in the current date in MM/DD/YY format. I.E. if it is Nov. 19

th

, 2011 we’ll type in 111911.

18.Utilize the number keys again to input the current date in the format listed, You’ll see that

date show up in the bottom screen in the following format: 11 19 2011

Содержание Baxter 6201

Страница 1: ...e to tug tubing This will ensure an occlusion was not created when closing the door and will also make sure the tubing isn t being pinched at all by the clip Tip 3 To Avoid Nuisance Occlusion Alarms If you ve been using the tubing for a while try using a fresh section of the tubing by moving the clip down and loading this new section of tubing Tip 4 To Avoid Nuisance Occlusion Alarms Occasionally ...

Страница 2: ...erved close regulating clamp recheck I V set loading and verify that the proper administration set is being used b If flow is again observed do not use the pump Have it inspected by service personnel Note Always close the administration set regulating clamp s before opening pump door and removing set 8 Attach set to I V access site 9 Turn pump on by pressing the ON OFF CHARGE key Verify that the p...

Страница 3: ...set that error code quickly and easily with the process detailed on pages 5 6 of this document so you won t have to send in your pump In some cases the F49 or F94 error code is caused by a bad battery If after completing this process you believe that this is the case please contact us so we may help you verify this and if so we will assist you with returning your pump for correction You can reach ...

Страница 4: ... can be adjusted to suit the needs of the hospital 4 Air sensor Detects air bubbles in the tubing The sensitivity is adjustable via the pump s configuration 5 Safety clamp Prevents accidental fluid flow by automatically occluding the tubing whenever the pump door is opened Provides Free Flow Protection 6 Slide clamp slot When the slide clamp loading option is enabled the I V set s slide clamp must...

Страница 5: ...s Tot Vol Status and has the text Next lit above it 9 This key will be used to cycle through the screens of the Modify Config mode 10 Press this key approx 18 times until you see the top screen say Time with some numbers below it 11 Don t worry about what time the pump is currently saying we are now going to program in the current time at your location in 24 hour or Military format I E if it s 2 3...

Страница 6: ... mode and be ready for use or it may come up with an additional Common Failure code of F 91 22 F 91 error is a common occurrence after this process and simply needs to have the pump powered off then back on again using the On Off key 23 Your pump should now be ready for use Although they can be caused by other factors typically these errors are caused by the main battery of the pump being complete...

Отзывы: