2 / 2
P/N 3102365-
EN • REV 003 • ISS 14FEB18
To perform an initial installation test:
1. Visually inspect each detector and verify that it is installed in the
correct location. Make sure it is not adversely affected by factors
not apparent on the plans.
2. Remove the detector from its base and verify that the proper
detector address, trouble signals, and messages are reported.
3. If wired for Class A operation, verify that the detector continues to
operate first with SLC_IN disconnected, and then with SLC_OUT
disconnected. (Refer to the installation sheet for the base.)
4. Place a momentary ground fault on the SLC to verify operation of
ground fault detection circuitry.
5. Perform a function test, as described below.
To perform a function test:
1. If desired, use the fire alarm control panel to put the detector or
zone into a service group for testing. (Refer to the panel technical
reference manual for instructions.)
2. Activate the heat sensor using a hair dryer (maintaining a distance
of three inches) or using a Testifire detector tester per the
manufacturer’s instructions.
Maintenance
To ensure proper operation, plan maintenance in accordance with the
requirements of the authority having jurisdiction. Refer to
CAN/ULC-S536
Standard for the Inspection and Testing of Fire Alarm
Systems
and NFPA
72 National Fire Alarm and Signaling Code.
Refer to application bulletin P/N 3102403-EN for additional information
and cleaning instructions.
Specifications
Operating voltage
15.20 to 19.95 VDC
Current
Normal operating
Alarm
51 µA
68 µA
Vibration level
10 to 35 Hz, with an amplitude of
0.01 in.
Maximum spacing [1]
50 ft. (15.2 m) centers
Compatible bases
Standard
Relay
Isolator
Audible
SIGA-SB, SIGA-SB4
SIGA-RB, SIGA-RB4
SIGA-IB, SIGA-IB4
SIGA-AB4, SIGA-AB4G
SIGA-AB4GT, SIGA-AB4G-LF
Compatible detector testers [2]
Testifire 2000
Operating environment
Temperature
Relative humidity
32 to 100°F (0 to 38°C)
0 to 93% noncondensing
Storage temperature
−4 to 140°F (−20 to 60°C)
[1] When replacing SIGA-HRS and SIGA-HFS detectors with the
SIGA-HFD, ensure that the spacing is 50 ft. (15.2 m) or less.
[2] Requires the SIGA2-TSTSPACER Testifire adapter assembly.
Regulatory information
North American
standards
CAN/ULC-S530-M91, UL 521
UL and ULC fixed-
temp alarm rating
135°F (57.2°C)
Actual alarm point
129 to 141°F (53.9 to 60.6°C)
Response time
index
Quick(evaluated by FM Approvals)
FCC compliance
This device complies with part 15 of the FCC
Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any
interference received, including interference that
may cause undesired operation.
Industry Canada
compliance
This Class A digital apparatus complies with
Canadian ICES-003.
Contact information
For contact information, see www.edwardsfiresafety.com