1.4 SPECIFICATIONS
Table 1-1
MODEL
1100
1100RS
Wavelength Range
335-1000nm
335-1000nm
Spectral Slit Width
20nm 10nm
Wavelength Accuracy
±
2nm
±
2nm
Wavelength Readability
2nm
Stray Radiant Energy
≤
0.5%T@340nm &400nm
≤
0.5%T@340nm & 400nm
Transmittance
0%T to 125%T
0%T to 125.0%T
Absorbancy
0A to 2A
0A to 2A
Photometric Accuracy
±
2.0%T
±
1.0%T
Photometric Noise Level
±
2.0%T
±
1%T
Power Requirement
115-volt, 60Hz, or 230-volt, 50Hz
Size
408mm(W)
×
308mm(D)
×
180mm(H)
Weight
5.5kg
Specifications subject to change without notice