TEMP. AND DIFF. PRESS. INDICATOR

3

2.  Mount the frame in the Surface cutout.

1.  Cutout required in the back wall or partition : 125[H] X 100[W]

3.  Fix the frame on surface by screws.

4.  Insert the Power & Communication Cables.

100

125

100

145

155

110

100

145

155

110

W

all OR P

artition

surf

ace

FRAME

Indic

at

or

3.  Mount the Indication in the Frame.

4.  Terminate the Supply and Comm. wires to Indicator.

7.  Fix the Indicator on Frame by screws.

2.2     Mounting

Cut Out

Содержание UT-1405

Страница 1: ...sales utplindia in 91 7046223333 9427301436 www utplindia in www Model No UT 1405 TEMP AND DIFF PRESS INDICATOR CLEAN ROOM INDICATOR INSTRUCTION MANUAL M UT 1405 UT 1405 888 8 5 0 2 8888 0 3...

Страница 2: ...AGE Power Supply Wiring Requirements Applying 240V AC to a controller rated for 12 24VDC will severely damage the controller and is a fire and smoke hazard This controller is suitable for connection t...

Страница 3: ...CHART 4 1 CONFIGURATION MODE 4 OPERATIONAL MODES 4 2 SETUP MODE 4 3 CALIBRATION MODE 4 4 CHANGE PASSWORD MODE 4 5 BUZZER MODE 5 COMMUNICATION MODBUS RTU PROTOCOL 5 1 INTRODUCTION 5 2 MODBUS CONFIGURAT...

Страница 4: ...INSTALLATION Indicator Frame NOTE All Dimensions are in mm The UNITECH s Clean Room Indicator T DP UT 1405 has been specially developed keeping process measurement requirements of Clean rooms paramet...

Страница 5: ...ired in the back wall or partition 125 H X 100 W 3 Fix the frame on surface by screws 4 Insert the Power Communication Cables 100 125 100 145 155 110 100 145 155 110 W all O R Partition surface FRAM E...

Страница 6: ...w scale Full range as per input selection General Specificaion Power Supply 24VDC Power Consumption 4 Watts Maximum Prog parameters Modbus RTU RS 485 Isolated Communication Baud rate Parity Node addre...

Страница 7: ...meproof model available for Hazardous applications This instrument enables you to measure parameters in the following fields HVAC System Heating ventilation Air conditioning cooling Pharmaceutical ind...

Страница 8: ...in Programming Mode NORMAL VIEWS ENTER KEY ENTER Key allows enter into the Parameter moving from one parameter to another and accept saving a new value of parameter or a status of parameter UP KEY UP...

Страница 9: ...bUZZ 1SP1 1SP2 1SP3 1SP4 2SP1 2SP2 2SP3 2SP4 Password Protected CAL CAL1 CAL RANG WARNING INJURY Programming parameter should be accessed only by Personnel who are technically competent to do so Local...

Страница 10: ...TEMP Alarm 1 Type 1sp1 Default type should be Lo Values are Low Low Low High High High LL LO H1GH HH II Input 1 TEMP Alarm 2 Type 1sp2 Values are Low Low Low High High High LL LO H1GH HH III Input 1 T...

Страница 11: ...RM STATUS ENABLE DISABLE aLST Enable OR Disable Alarm of particular Input Press key to access sub parameters the sub parameters are I Input 1 TEMP Alarm 1 1sp1 Values are Enable Disable En D1s Values...

Страница 12: ...ETER COMM Device Active Master Slave Communication configuration parameter Press key to access sub parameters the sub parameters are Device Address for Communication 6 INPUT DAMPING daMp Damping param...

Страница 13: ...the sub parameters are III Hours 24Hr hour Values are between to 0000 0024 Real time Hours Present Calender Date IV Day Date day Values are between to 0000 0031 Present Month IV Month MOn Values are b...

Страница 14: ...DP Alarm 2 Set Point 2sp2 VII Input 2 DP Alarm 3 Set Point 2sp3 VIII Input 2 DP Alarm 4 Set Point 2sp4 V Input 2 DP Alarm 1 Set Point 2sp1 2 Input 1 TEMP Alarm 1 Set Point 1sp1 Set any value between...

Страница 15: ...rd for access Setup mode 1234 II Input 1 TEMP High Calibration 1p1s Apply to High Span signal to Input 1 TEMP and press key IV Input 2 DP High Calibration 1p2s Apply to High Span signal to Input 2 DP...

Страница 16: ...sword 1234 Set the new Password for Buzzer Mode BUZZ 5 BUZZER MODE PASSWORD BUZZ Change Password Mode default password 1234 Set the new Password for Change Password Mode pASC 6 CHANGE PASSWORD MODE PA...

Страница 17: ...405 Device 2 UT 1405 Device 3 UT 1405 Device 256 2 core 1 sq mm shielded cable COMPUTER SCADA TEMP ALARM 2 SET POINT TEMP ALARM 3 SET POINT TEMP ALARM 4 SET POINT 40007 40009 40008 40010 40011 40012 4...

Страница 18: ......

Отзывы: