SPECIFICATIONS
S
ULFI
L
OGGER
™
SENSORS
SulfiLogger™
S1/X1-1020-5mgL S1/X1-1120-5mg S1/X1-1120-1000ppm X1-1220-1000ppm
Environment
Wastewater
Pressurized
wastewater
Pressurized
gases
Pressurized
gases
Measurement
range
0-5 mg/L
0-5 mg/L
0-1000 ppm
0-1000 ppm
Water depth/
Pressure
10m
3 barg
3 barg
10 barg
Input power
20 to 30 V
20 to 30 V
20 to 30 V
20 to 30 V
Output data
RS-232 or
4-20 mA
RS-232 or
4-20 mA
RS-232 or
4-20 mA
RS-232 or
4-20 mA
Mounting
Free hanging
Wall mounted
In-line
Flow cell
In-line
Flow cell
In-line
Flow cell
T
ECHNICAL SPECIFICATIONS
Accuracy
+/- 5% of full range
Response time (t90)
<25 seconds
Detection limit
1% of full range
Operating humidity range
0-100%
Design temperature range
-20
◦
C to +60
◦
C ( -4
◦
F to +140
◦
F)
Operating temperature range
0
◦
C to +40
◦
C ( 32
◦
F to +104
◦
F)
M
ECHANICAL SPECIFICATIONS
Length
240 mm
Diameter
Ø48.3mm / 1 ½” pipe
Weight
0.85 kg
Material
Stainless Steel - EN 1.4404 (316L)
Warranty
3 months
C
HEMICAL INTERFERENCES SPECIFICATIONS
Gas
Formula
Interference in gas (%)
1
Interference in liquid (%)
2
Methane
CH
4
0
0
Carbon dioxide
CO
2
0
0
Nitrogen
N
2
0
0
Oxygen
O
2
0
0
Air
O
2
, N
2
, Ar
0
0
Nitrous oxide
N
2
O
0
0
Ammonia
NH
3
0
0
Hydrogen
H
2
0.8
96
Carbon monoxide
CO
0.6
77
Dimethyl sulfide
(CH
3
)
2
S
18
18
Methyl mercaptan
CH
3
SH
174
44
Ethyl mercaptan
C
2
H
6
S
13
14
Sulfur dioxide
SO
2
40
1
1
Given as signal for the interfering species in % of H