39
5
Function Description Related to Indicated Values
39
Function Description Relate
d to Indicated Values
Chapter
5
5-3.
Moving average filter
This filter takes the moving averages of converted data and reduces fluctuation of indicated values.
The fluctuation of indicated values can be reduced further by performing this function more number
of times; however, response to input becomes slower. Torque and rotation speed can be individually
set.
<Setting value>
OFF (0), 2 to 300 [times]
* One time represents one sampling of A/D convert.
■
Execution procedure
1.
Switch the PC application screen to
the "Settings" tab.
2.
Input values in the moving average filter
(torque)/moving average filter (rotation
speed) respectively.
3.
When writing setting values to the
TM201, click the "Set" button.
4.
When the "Set" button is clicked, a
confirmation dialog to overwrite is
displayed, and click "Yes (Y)".