30
4
Calibration
30
Calibrat
ion
Chapter
4
(2) Unit (torque) setting
The unit of torque for calibration is selected.
<Setting value>
mNm, Ncm, Nm, kNm, kgm (kgfm),
kgcm (kgfcm), gcm (gfcm)
(3) Min. scale division (torque) setting (optional if there is no change.)
Min. scale division of the torque (scale interval, scale
division) is selected.
<Setting value>
1, 2, 5, 10, 20, 50, 100
(4) Calibration setting is written to the TM201
The values selected in (1) to (3) are written to the TM201 in the following procedures.
■
Execution procedure
1.
Click the "Set" button.
2.
When the "Set" button is clicked, a
confirmation dialog to overwrite is
displayed, and click "Yes (Y)".
(5) Zero calibration
Zero point is registered while torque load is not applied to UTM
Ⅱ
. Furthermore, if the initial
torque remains in the jig and so on for the purpose of measurement, the zero point may be
registered while the initial torque is being applied; however, the initial torque range needs to be
subtracted from the signal input range.
■
Execution procedure
1.
Select "Zero" for calibration mode.
2.
Click the "Execute" button.
3.
When the "Execute" button is
clicked, a confirmation dialog to
overwrite is displayed, and click "Yes
(Y)".