Moving average filter (setting mode 2-2)
This function restrains the indicated value from fluctuating by moving-averaging the A/D-
converted data.
The moving average times can be selected in the range of OFF (1 time) - 512 times. With an
increasing number of moving average times, the indicated value becomes more stable, while the
response becomes slower. On the other hand, with a decreasing number of moving average times,
the response becomes faster, while the indicated value becomes easier to fluctuate.
Set an optimum value according to the type of weighing.
5-5.
Filter in stable condition (setting mode 1-5)
This function automatically inserts the digital filter to restrain the indicated value in a stable
condition from fluctuating. Select whether or not to insert it.
Definition of stability is given by "5-6.Motion detection(MD) (setting mode 1-5, 2-3)" on page 33.
Load cell
Moving
F701-P
OFF
Setting times
Stable
average filter
Digital
filter
(fixed at 256 times)
When Stable is OFF or
the filter in stable condition
(
)
A/D
converter
Digital
low pass filter
Setting frequency
(
)
is not inserted.
Weighing function 2
Filter in stable condition
0: Insert (256 times)
1: Not insert