Chapter
4
4-4.
Zero calibration
Register the initial zero point.
- Check around the load cell (scale) for unwanted loads such that foreign matter is placed or there
is contact with peripheral equipment.
- Check that "
" is on.
(When the indicated value is not stable, accurate calibration cannot be performed.)
1.
Select setting mode 9 (calibration mode).
2.
Register the zero point.
3.
Zero calibration ends with the weight value display of "0".
* If a calibration error is displayed, take measures according to the error message, and perform zero
calibration again. (See "11-2.Over scale/error display" on page 100.)
kg
Display of zero calibration in progress.
→
kg
Weight value
Setting mode No.