4.CALIBRATION
18
4-3. Calibration Setting
4-3-1. LOCK2 (Calibration value LOCK)
Set values relating to calibration are locked so as not to be changed by misoperation.
Cancel the lock before performing calibration.
Alternatives:
[
OFF
]
[
ON
]
4-3-2. Minimum scale division (Can be omitted if there is no change)
Set the minimum scale division of the indicated value.
Setting range: 001 to 100
4-3-3. Decimal place
Set the decimal point place of the indicated value. Selection can be made from the following.
Alternatives:
[
NONE
]
[
0.0
]
[
0.00
]
[
0.000
]
4-3-4. Zero calibration
Set the zero point in no-load condition.
4-3-5. Equivalent input calibration
Set the rated output value and reading of the sensor.
Rated output value:
0.001 to 3.000mV/V
Rated value:
00001 to 99999
MODE
Twice
[
OPERATION
]
[
LOCK2
]
→
→
→
MODE
[
CALIBRATION
]
[
MIN. SCALE DIV.
]
→
→
Once
→
MODE
[
CALIBRATION
]
[
DECIMAL PLACE
]
→
→
Once
→
→
Once
MODE
[
CALIBRATION
]
[
ZERO CAL.
]
→
→
Once
→
MODE
[
CALIBRATION
]
[
EQUIV. CAL.
]
→
→
Once
→
About the [BALANCE WGT VALUE] setting
For performing calibration, weight value input can be omitted by setting a weight
value to [BALANCE WGT VALUE] beforehand. [BALANCE WGT VALUE] is ganged
with the weight value inputs of [SPAN CAL.] and [EQUIV. CAL.].
Содержание F600AT
Страница 1: ...F600AT 01MAY2015REV 3 01 WEIGHING CONTROLLER OPERATION MANUAL...
Страница 81: ......