Error release
- Common to load errors and displacement errors
- Load errors
- Displacement error
Error
Error release
Perform zero calibration within the zero calibration range.
Perform actual load calibration within the actual load calibration range.
Or, perform equivalent input calibration within the equivalent input calibration
range.
Bring the sensor input into the signal input range/internal count range.
Load : Make the load indicated value -9999 to +9999.
Displacement : Make the displacement indicated value -9999 to +32000.
Error
Error release
After the indicated value falls within the set Overload value, turn OFF
→
ON the
Reset input, or press
on the Main screen.
Turn OFF
→
ON the Reset input, or press
on the Main screen. Or, turn
the power off and then on again.
Error
Error release
Turn OFF
→
ON the Reset input, or press
on the Main screen. Or, turn
the power off and then on again.