5.

 

Circuit description 

 

5.1

 

Power supply 

 
The device is mains powered. It sports an all-range off-line switched mode power supply , 
meaning it can be used all over the world without having to select a working voltage or 
frequency. But of course the the plug of the power cord has to fit into the local outlet type. 
 

5.2

 

Patient Current Routing 

 
The electrotherapy outputs of the main unit are connected to the vacuum unit, at the rear, and 
the electrodes are connected to the front of the vauum unit. As long as the vacuum is not used 
the traditional  electrodes function as before, even when the vacuum unit is not switched on. 
As soon as the vacuum is turned on the patient current is switched from the traditional 
electrodes to the connectors for the hoses at the other end of which the suction cups are. In the 
hoses is a wire that feeds the patient current to a metal plate inside the cup. Contact with the 
the skin of the patient is established by a moisturized sponge. The switchover always happens 
for channel A but the operator can add channel B by pressing a button that toggles the choice 
between traditional and vacuum electrodes. 
 

5.3

 

Vacuum 

 
When the vacuum control is turned up, air is pumped out  to create the vacuum until the set 
level is reached. The hose connectors limit the air flow into each hose so one can put the 
vacuum electrodes one at a time in their place. When channel B is not used, its connectors 
should be plugged by connecting them to each other by a hose, otherwise the pump keeps 
running and deeper vacuum levels cannot be reached. 
 

5.3.1

 

Control Loop 

 
A pressure sensor measures the vacuum level and a control loop powers the pump if the set 
level is deeper or opens a valve to let air in when lower . In continuous mode this results in only 
the pump being run very lightly, making up up for the small air leakage between the cups and 
the skin and at the connectors. 
 

5.3.2

 

Pulsed Vacuum 

 
When the electrodes are in place the operator can pulsate the vacuum to mitigate the effect 
vacuum has on patients with a vulnerable skin. When pulsating, the vacuum is raised by 50% 
during 1.5 second and then returns for 1.5 to 4.5 seconds to the level set for continuous 
operation. As a result the pump will run during the 1.5 second period of deepened vacuum after 
which the valve will be opened until the base level is reached again. 
 

Содержание Phyaction V

Страница 1: ...pply 5 2 6 Patient Current Routing 5 3 6 Vacuum 5 3 1 6 Control Loop 5 3 2 6 Pulsed Vacuum 5 3 3 7 End of Treatment 5 3 4 7 Water management 6 8 Safety inspection list yearly check up 7 9 Hardware lab...

Страница 2: ...r of the Uniphy family it is built according to the requirements of the MDD security standards All functions of the unit are controlled by a microprocessor ensuring a high degree of reliability and sa...

Страница 3: ...e able to provide the following data for each unit Instrument data Part and serial number Gymna has the original configuration of each unit The Phyaction V has following critical parts Ref 324597 Main...

Страница 4: ...er name Customer address Instrument name Serial number Date of delivery Service activities Date Error complaint Service action Critical spare parts exchange list Critical part name Old serial number N...

Страница 5: ...socket required Insulation Type BF floating patient circuit Fuses 2 x T2AL250V Volume water reservoir 180 ml Working pressure continuous vacuum 38 320 hPa Working pressure pulsation vacuum 58 480 hPa...

Страница 6: ...el B by pressing a button that toggles the choice between traditional and vacuum electrodes 5 3 Vacuum When the vacuum control is turned up air is pumped out to create the vacuum until the set level i...

Страница 7: ...avoidably droplets are sucked into the system These droplets do not only contain water but also fibers from the sponges flakes of skin and other substances present under the cups To prevent these to r...

Страница 8: ...the keys and the display are undamaged Mains entry and power cord are undamaged The output connectors are undamaged The electrode connectors and cables are undamaged All the operation functions work T...

Страница 9: ...ut The last 3 alphanumerical numbers will begin with A A A to Z Increase in case of a principle technology change Components are not anymore compatible with the previous layout All previous changes ma...

Страница 10: ...ing static sensitive devices such as PC boards of the Phyaciton V the following precautions should be observed Persons should be earthed by means of a wrist strap Ground all electrical equipment work...

Страница 11: ...pply PCB Unscrew the 4 screws of the power supply PCB and replace with a new PCB 8 3 Replacement of the Main PCB The main PCB is one board in production that is divided over the front of the unit To r...

Страница 12: ...m the patient We have noticed that is some cases the vacuum is switched off even when there is current detected In order to solve this the unit needs to be modified Step 1 open the unit Step 2 remove...

Страница 13: ...324597 MAIN PCB PHYACTION V VACO 200 324650 CABLE ASSY ET IN VACO500 PHYACTION V 324727 CABLE ASSY MAINS INLET SMPS 200 SERIES 324793 BOTTOM BOX PHYACTION V 324804 METAL TOP COVER PHYACTION V 324837 C...

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ...R247 ELVA003 LO200 V21 pcb Tue Dec 13 10 38 58 2005...

Страница 18: ......

Отзывы: