Service Information Phyaction 740 & 790-series
Page 36
4. Peripherals (during normal operation)
40 ADC_
TIMEOUT
ADC time out
41 LCD_
11*
TIMEOUT
LCD time out
42
OC_ERR
Optocoupler error
OCdet (ADC)
43 I2C_
TIMEOUT
I²C time out
(0=send,1=receive)
keyboard (KB)
front panel (FP)
ultrasound (US)
dose (DS)
0/1
40h
46h
4Ah
96h
44 KEYBOARD_
ERR
Keyboard error
0
sent / received
45 BININ_ERR Instable
digital
signal 0
number
low/high
func-
tion
49 INT_ILLEGAL Illegal
interrupt
vector
5. and 6. Safety circuit current generator
50
POTMET_ERR Dose regulator not in zero position
51 ERRORMF_
ERR
MF-error line active
53
POWER_ERR
Supply voltage sum error
Vsum (ADC)
54
POTADC_ERR Dose regulator voltage not 0 V
VPOTmf (ADC)
55
ERRACT_ERR MF-error line cannot be deactivated
56 POTGUARD_
ERR
Dose regulator guard error
57 NOERRORMF
_ERR
MF-error line inactive at poweron
58
VTEMP_ERR
Temperature much too high
Vtemp (ADC)
59
MFPOT_ERR
Switch of intensity regulator
defective
0 VPOTmf
(ADC)
60
UI_ERR
V/C converter error
Vmf (ADC)
61
RELPOS_ERR
Safety relay in wrong position
62 RELSWITCH_
ERR
Safety relay cannot switch
Vmf (ADC)
63
PAT_ERR
Patient connected
Vmf (ADC)
Содержание Phyaction 740 Series
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 58: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......