INSTRUCTION MANUAL FOR UNIPHOS PRECISION AIR SAMPLING PUMP MODEL: 

ASP-40  

 

 

The UNIPHOS precision gas detector consists of two parts viz. i) the gas detector tubes and ii) air 

sampling  pump  (ASP-40).  Together  they  form  the  UNIPHOS  gas  detection  system.  The  air 
sampling pump is used to draw a precisely defined volume of sample air through the detector tube 
for  determining  the impurity  gas  concentration  in  the  air sample. This manual  describes  the  air 
sampling pump and its operation for impurity gas measurement using detector tubes.  

 

 

 

 

1.  DESCRIPTION  

  
ASP-40 is a piston and barrel air-sampling pump (Fig.-1). For measurement, a detector tube with 

both ends cut is connected to the rubber tube connector at the end of the barrel. When the piston 
is withdrawn by pulling the handle, sample air is drawn into the barrel through the detector tube.  

 

  

  

Fig.-1 Uniphos Precision Air Sampling Pump (ASP-40)  

  
  
  
  
  
  

  

  
  

Fig.-1 shows the different parts of the air-sampling pump, which are described in the following 

section.  

  

1.1 

Tip cutter:  

The tip cutter is provided on the pump to break off both ends of the detector tube before it is 

inserted  into  the  rubber  tube  connector  for  drawing  the  air  sample.  The  broken  tips  are 
collected into a storage space which when filled is to be emptied by pulling off the rubber cap 
on the opposite side.  

  

Содержание ASP-40

Страница 1: ...UNI AIR IN PHOS PRECISION SAMPLING PUMP MODEL ASP 40 STRUCTION MANUAL...

Страница 2: ...is a piston and barrel air sampling pump Fig 1 For measurement a detector tube with both ends cut is connected to the rubber tube connector at the end of the barrel When the piston is withdrawn by pu...

Страница 3: ...nd reaching 50 cc 1 2 stroke or 100 cc full stroke mark For locking the red dots on the stopper and shaft have to be aligned In this position when the shaft is pulled it gets locked at 50 or 100 cc ma...

Страница 4: ...he sample air and produces a coloured stain in the sensing chemical The length of the coloured stain measures the impurity gas concentration in ppm on the tube scale The measurement involves the follo...

Страница 5: ...p the end of the tube in the sample gas source while the gas is being drawn through the tube for the allotted sampling time approximately 1 3 minutes as indicated on the tube sheet The sampling is con...

Страница 6: ...The following steps are recommended 3 1 Insert a new detector tube without breaking its ends into the rubber tube connector 3 2 Align the red dots and pull the pump handle a full stroke 3 3 Wait for t...

Страница 7: ...bber packing of the piston iv Inspect the piston gasket on the end of the plunger for damage To replace the gasket use a small screw driver or other sharp tool to pull off the worn gasket and slip a n...

Страница 8: ...d and if necessary replaced or corrected Symptom Corrective action Rubber tube connector is loose Rubber tube connector is worn or damaged Piston sticks or slip too easily One way valve s on the pump...

Отзывы: