27 of 33
Trouble shooting guide:
Normal fault
Cause
Remedy
Transmitter unit
not getting
powered on
Loose connection
on supply cables
or
Reverse polarity
Verify the cable
connections on
Connector CN7 of the
bottom board.
Transmitter
LCD/OLED showing
Error-
No sensor
sensor is not
properly plugged
in the sensor
housing
Remove the sensor &
re-plug it again
carefully
Transmitter
LCD/OLED showing
Error-
Sensor
–
drift
Or Over-Range
Sensor due for
calibration
Go for calibration
Transmitter is
initialized
properly but loop
current 4- 20 mA
signal is not
available.
Signal out wire
(Green) shorted
with the ground
wire.
Cross verify the
wiring.
Check “4mA
Enable/Disable” option
Содержание 500DT
Страница 9: ...9 of 33 Figure 2 Outline drawing Front side views with outline dimensions...