Transport, setup and acceptance
22
5.2 Ambient conditions
ATTENTION
Ambient conditions during storage and set up.
Observe the stipulated ambient conditions. Contact Union Instruments GmbH if
the process gas analyser has been stored for more than three months or needs to
be operated or stored under ambient conditions other than those specified.
5.2.1 Storage conditions
If condensation water freezes in the process gas analyser, this can cause defects.
Protect the process gas analyser against frost when putting it into storage.
Ambient temperature:
-20 - 60°C
Humidity:
0 - 95% relative humidity
Ambient pressure:
700 - 1400 hPa (0.7 - 1.4 bar)
5.3 Set up and connection
5.4 Setup site
The setup site for the process gas analyser must satisfy the following conditions:
•
Clean dry room (with the exception of INCA5000/INCA6000)
•
No direct exposure to sun
•
Insure a clean, sufficient amount of ambient air for undistorted measurements
•
Ensure that the loadbearing capacity of the wall is sufficient
WARNING
Leaking process gas can pose a hazard and needs to be discharged by the
operator into a safe environment.
Содержание INCA6003
Страница 1: ...Apr 2016 V1 10 Translation of the original operating instructions Process gas analyser INCA6003 ...
Страница 3: ...3 Dimensions ...
Страница 10: ...EU Declaration of Conformity 10 ...
Страница 34: ...Startup switching on 34 ...
Страница 48: ...Operation 48 ...
Страница 50: ...Decommissioning switching off 50 ...
Страница 54: ...Maintenance 54 ...
Страница 58: ...Service 58 ...
Страница 60: ...Associated documents 60 ...
Страница 62: ...Disposal 62 ...
Страница 64: ...Spare parts 64 ...
Страница 65: ...65 16 Annex ...
Страница 66: ...Annex 66 EU Declaration of Conformity Flame arrester ...
Страница 67: ...Annex 67 ...