UNION Instruments GmbH
INCA Analyzer Series
– Instruction Manual
Output option
Page 43 of 59
4
Change channel / measurement point
5
Calibrate with purge gas
6
Calibrate with calibration gas I
7
Calibrate with calibration gas II
15
Error
secondsInState
4
Time in seconds how long analyzer has been in current state (
measureState).
dataValidDiscontChnls
1
Data validity of values in
currValues[10]
for discontinuously
measured channels (e.g. H
2
S oder H
2
). Only valid for devices
which are configured as continuous measurement devices, but
also have channels measuring discontinuously.
Value
Definition
0
Values are not valid
1
Values are valid
gasCoolerTemp
2
Temperature of gas cooler
(if information is available, otherwise 65535 (or 0xFFFF) is
returned)
Value is transmitted with decimal place information (DPI)
Word:
T_COOL DPI: yes (2)
Unit: °C
irTemp
2
Temperature of infrared cuvette
(if information is available, otherwise 65535 (or 0xFFFF) is
returned)
Value is transmitted with decimal place information (DPI)
Word:
T_IR
DPI: yes (2)
Unit: °C
paroxState
2
State of Parox sensor
(if information is available, otherwise 65535 (or 0xFFFF) is
returned)
Value
Definition
0x0000
OK
0x0400
Parox sensor is warming up
outerCaseTemp
2
Ambient temperature (outer case)
(check if value is in valid range: e.g. -25.00 °C – 75.00 °C)
Value is transmitted with decimal place information (DPI)
Word:
T_OUT DPI: yes (2)
Unit: °C
useValidFlagDiscontChnls
1
Flag, which tells, if the ‘dataValidDiscontChnls‘ information is
supposed to be evaluated or not
reserve[156]
156
- nicht definiert -
Table 6: Data description