Technical data
9
1.10 Environmental conditions
NOTICE
When the combustion calorimeter is used outside of the environmental
conditions, additional measures (air conditioning of the combustion calorimeter,
etc.) must be agreed on with UNION Instruments GmbH!
Installation location:
Installation room required (
Chapter 4
)
Room temperature:
5 – 40 °C
Temperature change
≤ 5 °C per hour
Air humidity:
0 – 95 % rel. humidity
Ambient pressure:
800 – 1100 hPa (0.8 – 1.1 bar)