Technical data
8
1.8 Process gas
NOTE
The process gas must be free of condensate, dust and other contaminants, such
as naphthalene.
Gas connection inlet pressure:
20 - 40 mbar
Gas consumption:
30 - 40 l/h (min. rel. density 0.50 with 0.55 mm
nozzle)
1.9 Linearity and measuring ranges
The measuring ranges cannot be utilized from 0% to 100%. The range is depend-
ent on the gas composition. Ranges of 45-100 % are typical for a measuring range.
Hydrogen content in the gas increases the measuring range span. Inert gases such
as N
2
, O
2
, or CO
2
, reduce the measuring range span.
CWD device
series
CWD2005
CWD2005
CT
CWD2005
PLUS
CWD2005
DPC
CWD2005
SPC
CWD2000
Ex
Flare gas
0 – 15
3.0
––
––
––
––
0 – 15
0 – 15
––
––
Blast furnace
gas
3.5 – 6
3.0
––
––
3.5 – 6
3.0
3.5 – 6
3.0
3.5 – 6
3.0
––
––
Converter gas
4.5 – 9
1.5
––
––
––
––
4.5 – 9
1.5
4.5 – 9
1.5
––
––
Mixed gas
5 – 10
2.0
––
––
5 – 10
2.0
5 – 10
2.0
5 – 10
2.0
––
––
Coke oven gas
15 – 30
1.5
––
––
15 – 30
1.5
15 – 30
1.5
15 – 30
1.5
––
––
Biogas
25 – 35
1.5
––
––
25 – 35
1.5
25 – 35
1.5
25 – 35
1.5
––
––
Natural gas
25 – 48
1.5
30 – 47
1.0
25 – 48
1.0
25 – 48
1.0
25 – 48
1.0
25 – 47
1.0
Refinery gas
25 – 50
1.5
––
––
25 – 50
1.5
25 – 50
1.5
25 – 50
1.5
––
––
Liquefied petro-
leum gas (LPG)
40 - 90
1.5
––
––
40 - 90
1.5
40 - 90
1.5
40 - 90
1.5
––
––
2.0
2.0
Measuring range [MJ/m] / Accuracy [± MBE]
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...