23
24
TROUBLESHOOTING
ERROR code 3
(E3)
appears
The unit does not measure
Check whether tube connection of
the cuff is secured to the unit
properly
ERROR code 4
(E4)
appears
The cuff may not be applied
If you have a very weak or irregular
heart beat, the device may have
difficultly in determining your blood
pressure
There is a measuring Error
Sit comfortably and still. Fasten the
cuff again carefully
ERROR code 5
(E5)
appears
Cuff over inflated
The measurement range is over
300 mmHg. It is recommended to
see the doctor as soon as possible.
The monitor keeps
reinflating
Circuit locked
Remove and reinsert the batteries
and then proceed to take
measurement again.
ERROR code 6
(E6)
appears
Low battery
The battery power is too low to
function. Replace the batteries with
new ones.
TECHNICAL SPECIFICATION
Display
: LCD Display
Measurement Range
: Systolic Pressure: 50-250 mmHg ; Diastolic Pressure: 30-200 mmHg
Pulse : 40-180 beats/minute
Accuracy
: Pressure : +/-3 mmHg
Pulse : +/-5% of reading
Resolution
: Pressure : 1 mmHg
Pulse : 1 beat / minute
Measurement Method
: Non-invasive, Oscillometric method
Power Source
: 4 x 1.5V AAA batteries (battery life 300 times)
Optional AC Adaptor
: Input 100-240V, 50-60 Hz ; Output 6V@600mA (service life 5000 times)
Operating Temperature / Humidity
: +5˚C to +40˚C, 15-93% RH maximum
Storage Temperature / Humidity
: -25˚C to +70˚C, up to 93% RH maximum
Operation, storage and
: 700hPa to 1060hPa
transport atmospheric pressure
Outer Dimensions
: Approx. 164 x 99 x 46mm
Arm Circumference
:
Accessories
: Cuff, User Manual, Storage Pouch
Classification
: Application part Type BF
Key to symbols
: Application part Type BF
: Class II equipment symbol
Содержание AM2305
Страница 5: ...7 8 PART IDENTIFICATION DISPLAY READINGS...
Страница 6: ...9 10 CLOCK MODE STANDBY MODE Clock Mode Standby Mode INSTALL CHANGE BATTERIES...
Страница 7: ...11 12 TO SET CLOCK...
Страница 9: ...15 16 TO MEASURE YOUR BLOOD PRESSURE PRESSURE BAR INDICATOR...
Страница 10: ...17 18 WHO CLASSIFICATION INDICATOR TO RECALL AVERAGE AND PREVIOUS MEASUREMENT DATA...