5.
Using the screwdriver, or similar, insert the tool into the slot on the SHUTTER FUSE cap, apply
slight inward pressure and rotate the cap 1/8 turn, counterclockwise. The fuse cap will pop out
slightly.
6.
Remove the fuse cap and fuse.
7.
Replace the fuse with a known good one. Shutter fuse is a 0.6A, 5 x 20 mm, time-lag (slo-blo)
type.
8.
Reverse the procedure to re-install the fuse cap, reconnect all cables and power the unit back up.
Initial Operation and Testing
1.
The
VMM-T1
is manually selectable to operate from 115 VAC or 230 VAC (50-60Hz).
PLEASE BE SURE THAT THE UNIT IS PROPERLY SET UP FOR THE LINE
VOLTAGE TO BE USED.
Refer to the “
Voltage Change
” section of this manual for further
information and procedures.
2.
Once the line cord has been attached to unit and connected to properly grounded wall receptacle,
the unit may be energized. See “
Start Up
” section.
3.
Insert the 7-pin male connector of shutter interconnect cable into 7- pin female receptacle at rear
of unit labeled SHUTTER.
4.
Connect the 7-pin female connector of the shutter interconnect cable to 7-pin male connector on
shutter to be driven.
5.
Place
VMM-T1
POWER switch to the ON “1” position, the POWER LED will illuminate.
6.
Place the N.O./N.C. switch into the N.O. position. The shutter will open and remain open until
the switch is returned to the N.C. position. The DRIVER ACTIVE LED will illuminate when
this switch is in the N.O. position.
7.
All
UNIBLITZ
drivers provide the circuitry necessary to support shutters equipped with the
Solid State Synchronization option. Simply plug the shutter interconnect cable into the driver. If
the shutter is equipped with the Solid State Synchronization option, the green LED labeled
SYNC ACTIVE will illuminate when the shutter is in the open position. In addition, the SYNC
Output BNC will change to the active state when the shutter is open. The absence of the solid
state synchronization option will inhibit the operation of the SYNC output and SYNC ACTIVE
LED. The remainder of the
VMM-T1
systems will not be affected.
8.
See specifications and operator controls for additional operational information concerning other
features of the V
MM-T1
9.
Should the shutter and/or controller not respond as described previously, be sure line cord is
installed into the receptacle and connections to the shutter are properly made. Note that the
shutter output is also fused. The fuse is located on the front panel and is a 0.6A, 5 x 20 mm,
WARNING: Be sure power switch is in the off position
before connecting the power cord to line. Attach line cord
to the unit first before plugging into the AC power source.
WARNING