VCM-D1 User Manual
13
Initial Operation and Testing
The
VCM-D1
will operate from 115VAC or 230VAC (50-60Hz), this is manually
selectable.
*
PLEASE BE SURE THAT THE UNIT IS PROPERLY SET UP FOR THE LINE
VOLTAGE TO BE USED.*
Once the line cord has been attached to unit and connected to properly grounded wall
receptacle, the unit may be operated.
*BE SURE POWER SWITCH IS IN THE OFF POSITION BEFORE
CONNECTING PLUG TO LINE. ATTACH LINE CORD TO THE UNIT FIRST
BEFORE PLUGGING INTO THE AC POWER SOURCE.*
See “
Start Up
” section. Insert the 7- pin male ODU type connector of shutter
interconnect cable into 7- pin female ODU type receptacle at rear of unit labeled SHUTTER.
Connect the 7- pin female Wire-Pro connector of shutter interconnect cable to 7- pin male Wire-
Pro connector on shutter to be driven.
Place POWER switch to the ON “1” position, the POWER LED will illuminate. Place the
N.O./N.C. switch into the N.O. position. The shutter will open and remain open until the switch is
returned to the N.C. position. The SHUTTER ACTIVE LED will illuminate when this switch is
in the N.O. position.
All
UNIBLITZ
drivers provide the circuitry necessary to support shutters equipped with
the solid state synchronization option. Simply plug the shutter interconnect cable into the driver.
If your shutter is equipped with this option the green LED, labeled SYNC ACTIVE, will illu-
minate when the shutter is in the open position. In addition, the SYNC output will change to the
active state when the shutter is open. The absence of the solid state synchronization option will
only inhibit the operation of the SYNC output and SYNC ACTIVE LED. The remainder of the
VCM-D1
systems will not be affected. In addition the SYNC system can be disabled by sliding
the “G” FUNCTION switch to the upper position.
See specifications and operator controls for additional operational information
concerning other systems of the
VCM-D1
.
Should the shutter and/or control not respond as described previously, be sure line cord is
installed into the receptacle and connections to the shutter are made properly. Note that the
shutter output is also fused (located at the front panel - .6A time-lag).
TURN OFF THE UNIT
AND REMOVE THE PLUG FROM THE AC SOURCE BEFORE CHECKING FOR
BLOWN FUSES
. Be advised, a visual inspection of a fuse is usually
NOT
an adequate test to
determine if a fuse failure has occurred. Use a DMM (Digital Multi-Meter) or equivalent test
device to determine fuse continuity. Also, particular shutter units respond to different minimum
pulse widths. For example, a standard VS25 shutter has a minimum exposure pulse of 6ms, if the
timing is set for an exposure pulse width less than 6ms, the shutter may not open fully. If the unit
still does not operate properly, please notify Vincent Associates immediately.
CAUTION
CAUTION
CAUTION
Содержание VCM-D1
Страница 1: ...User Manual UNIBLITZ VCM D1 Shutter Driver 14 0025 Version 2 0 2004...
Страница 4: ...VCM D1 User Manual 4...
Страница 17: ...VCM D1 User Manual 17 Notes...
Страница 18: ...VCM D1 User Manual 18 Figure 4 VCM D1 Rear Panel Operator Controls and I O...
Страница 30: ...VCM D1 User Manual 30 Figure 9 Overall VCM D1 Dimensions...