Measurement Operation(3)
Warning
Never attempt an in-circuit current measurement
where the open-circuit voltage between the circuit
and ground is greater than 60V.
If the fuse burns out during measurement, the Meter
may be damaged or the operator himself may be hurt.
Use proper terminals, function, and range for the
measurement. When the testing leads are connected
to the current terminals, do not parallel them across
any circuit.
The measurement ranges of AC current are: 200.0
µ
A,
2.000mA, 20.00mA, 200.0mA and 10.00A.
The measurement ranges of DC current are: 200.0
µ
A,
2.000mA, 20.00mA, 200.0mA and 10.00A.
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-
voltage capacitors.
2. Insert the red test lead into the
µ
A
mA
or
10A
terminal
and the black test lead into the
COM
terminal.
Use the 1
0A
terminal and
10A
or
10A
range if
the current value to be tested is an unknown.
3. Set the rotary switch to an appropriate measurement
position in
A
range or
A
.
4. AC current is displayed as an effective value of since
wave (mean value response).
5. Break the current path to be tested. Connect the red
test lead to the more positive side of the break and
the black test lead to the more negative side of the
break.
6. Turn on power to the circuit.
The measured value shows on the display.
Model UT90B: OPERATING MANUAL
Содержание UT90B
Страница 2: ...Model UT90B OPERATING MANUAL 2 ...
Страница 18: ......
Страница 19: ......
Страница 21: ...21 Measurement Operation 8 Model UT90B OPERATING MANUAL ...
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 27: ......
Страница 31: ...31 Model UT90B OPERATING MANUAL ...