28
X.Power supply and fuse replacement
1.Power supply setting
2.Fuse replacement
F2 F600mA 1000V
100V
240V
FUSE
AC power FUSE
AC power input socket
Power supply voltage setting
TO AVOID ELECTRIC SHOCK, REMOVE THE TEST
LEADS AND TURN OFF THE POWER INPUT BEFORE
OPENING THE CASE AND REPLACING THE FUSE.
POUR ÉVITER UN CHOC ÉLECTRIQUE, RETIREZ
LES CORDONS DE MESURE ET ÉTEIGNEZ
L'ALIMENTATION AVANT D'OUVRIR LA CASE
ET DE REMPLACER LE FUSIBLE.
SELECTOR
LINE(45-440Hz,28VA Max)
FUSE
100V
120V
AC 100V
AC 120/127V
220V
240V
AC 220/230V
AC 240V
AC 250V T250mA
AC 250V T125mA
SELECTOR
LINE(45-440Hz,28VA Max)
FUSE
100V
120V
AC 100V
AC 120/127V
220V
240V
AC 220/230V
AC 240V
AC 250V T250mA
AC 250V T125mA
1) Turn the red switch to the corresponding position
2) Setting steps:
a. Unplug the power cord
b. Turn the red switch to corresponding position
c. Selectable positions are shown below
1) Unplug the test leads from the instrument.
2) Turn off the power supply for the instrument
3) Open the fuse housing with a screwdriver.
4) Replaced the fuse with new one.
100V
120V
220V
240V
100V
120V/127V
220V/230V
240V
Position
Voltage
Demonstration
Description
Input
corresponding
voltage
1
2
3
4