49
Model UT81B: OPERATING MANUAL
l
The specification is same as Di Under Multimeter mode.
On the top right corner of the display shows the accuracy of the frequency which is around 2%. When the LCD
displays 1-10 period waveform, the reading is valid, otherwise the Meter displays “------ “
400
Ω
4k
Ω
40k
Ω
400k
Ω
4M
Ω
40M
Ω
250VDC or AC rms
Range Resolution Accuracy Overload Protection
E. Resistance
0.1
Ω
1
Ω
10
Ω
100
Ω
1k
Ω
10k
Ω
(1.2%+5)
(1%+5)
(1.2%+5)
(1.5%+5)
Range Resolution Overload Protection Remarks
F. Diode Test
A good silicon junction drops
between 0.5V and 0.8V.
250V DC or AC
1mV