Model UT60H: OPERATING MANUAL
18
Note
l
If the value to be measured is unknown, use the
maximum measurement position and reduce the
range step by step until a satisfactory reading is
obtained.
l
When the measured current is 5A, continuous
measurement is allowed.
l
When the measured current is between 5A-10A,
continuous measurement 10 seconds and interval
more than 15 minutes.
l
When DC current measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test and remove testing leads
away from the input terminals of the Meter.
Measurement Operation(4)
D. AC Current Measurement
(see figure 4)
Warning
Never attempt an in-circuit current measurement
where the open circuit voltage between terminals
and ground is greater than 60V DC or 30V rms.
If the fuse burns out during measurement, the Meter
may be damaged or the operator himself may be hurt.
Use proper terminals, function, and range for the
measurement.
When the testing leads are connected to the current
terminals, do not parallel them across any circuit.
Before the Meter is connected in serial to the in-circuit
being tested, turn off the power of the in-circuit..
Содержание UT60H
Страница 5: ...Model UT60H OPERATING MANUAL 5 ...
Страница 12: ...Functional Buttons ...
Страница 13: ...4 5 6 7 8 9 10 11 3 2 1 15 14 13 12 ...
Страница 14: ......
Страница 16: ......
Страница 21: ...Model UT60H OPERATING MANUAL 21 Measurement Operation 7 ...
Страница 32: ...Model UT60H OPERATING MANUAL 32 General Specifications ...
Страница 34: ...x x ...
Страница 35: ......
Страница 36: ... ...
Страница 37: ......
Страница 41: ...Model UT60H OPERATING MANUAL 41 ...