(figure 8)
Measurement Operation (6)
3.
Connect the test leads across with the object being
measured.
The measured value shows on the display.
Note
l
The test leads can add 0.1
to 0.2
of error to
resistance measurement. To obtain precision
readings in low-resistance measurement, that is the
range of 400.0
, short-circuit the input terminals
beforehand, using the relative value function button
REL
to automatically subtract the value measured
when the testing leads are short-circuited from the
reading.
l
For high-resistance measurement (>1M
), it is
normal taking several seconds to obtain a stable
reading.
l
If
reading with shorted test leads is not 0.5
,
check for loose test leads, incorrect function selection,
or enabled Data Hold function.
l
The LCD displays
indicating open-circuit for the
tested resistor or the resistor value is higher than
the maximum range of the Meter.
l
When resistance measurement has been completed,
disconnect the connection between the testing leads
and the circuit under test.
C. Capacitance Measurement
(see figure 8)
black
red
Model UT60A: OPERATING MANUAL
24
Содержание UT60A
Страница 9: ...The Meter Structure Model UT60A OPERATING MANUAL 9 ...
Страница 10: ...Model UT60A OPERATING MANUAL 10 ...
Страница 11: ...Model UT60A OPERATING MANUAL 11 ...
Страница 12: ...Model UT60A OPERATING MANUAL 12 ...
Страница 13: ...Model UT60A OPERATING MANUAL 13 ...
Страница 14: ...Display Symbols 1 Model UT60A OPERATING MANUAL 14 1 kM nFHz VA AC 15 2 3 4 5 6 7 8 9 10 14 16 ...
Страница 15: ...Display Symbols 2 Model UT60A OPERATING MANUAL 15 ...
Страница 16: ...Model UT60A OPERATING MANUAL 16 ...
Страница 43: ...Model UT60A OPERATING MANUAL 43 ...