background image

9

10

8

11

13

12

Button 

               Operation Performed

LEVEL Press  to  selection  of Auto  ranging,  Manual 

ranging. 

  The Meter is default to auto ranging.

  Press LEVEL button to switch to manual  

    ranging.  Press

button to toggle from 

    low to high range or from high to low 

    range.  Press and hold LEVEL button to 

    exit the manual ranging mode.  

FAST/

SLOW

  Press to select a FAST (125ms) or a  

    LOW (1 second) response time.  Select 

    FAST to capture noise peaks and noises 

    that occur very quickly.  Select the LOW 

    response to monitor a sound source that 

    has a consistent noise level or to aver

    age quickly changing levels.  Select LOW 

    response for most application. 

  Press and hold FAST/SLOW button to 

    enable display backlight.  Press and hold 

    FAST/SLOW button again to disable 

    display backlight.

  For UT352, press FAST/SLOW button to 

    store data:

      Press HOLD button to freeze data, the 

        LCD displays HOLD and M symbol and 

        the data being stored.  

      Press FAST/SLOW to store the data,   

        the  symbol RECORD and the number 

        of index blink for 0.5 seconds. 

      The Meter will automatically exit HOLD 

        mode.

      The Meter can store up to 63 data.

MAX/

MIN

  Press MAX/MIN to display the maxi

    mum or minimum reading.  The display 

    will update only when the measured value 

    exceeds the value presently in the display.

      Press the MAX/MIN button and the MAX 

        icon will display on the reading. The reading   

        displayed is the highest reading encountered  

        since the MAX mode was entered.

   Press and hold MAX/MIN button to 

     disable  or enable auto power off feature.  

     The Meter will automatically off  after 

     approximately 15 minutes of inactivity.

     The   icon display indicates that the 

     auto power off feature is active.

Button 

                Operation Performed

Calibration 

(see figure 4)

1.  Turn the Meter on.

2.  Put the Meter in the “A” weighting mode, FAST

     response mode, range set to 60~110dB, lock to MAX.

3.  Place the microphone onto the calibrator’s 1/2 inches 

     sound source hole.

4.  Turn the calibrator on, using 94dB@1kHz standard 

     sound source.

5.  Adjust the Meter’s CAL potentiometer located on the 

     side panel until the LCD displays 94.0dB

figure 4

Specifications

A.General Specifications

  Display: 3 1/2 digits, 1999 maximum 

  Overloading:

    Under range displays 

UNDER

    Over range displays 

OVER

  Battery Deficiency:  

    Change batteries as soon as 

 is displayed.

  Sampling Rate: 

    Fast Speed: 125 microseconds

    Slow Speed: 1 second

  Microphone: 1/2” electret condenser

  Drop Test: 1 meter pass

  Battery:  4 x 1.5V batteries (AA)

  Battery Life: Typical 20 hours continuous

  Dimension: 273 x 69 x 39mm

  Weight: around 386 g (including batteries)

B.Environmental Requirements

  For indoor use only.

  Altitude: 2000m

  Temperature and humidity:

       Operating:

         0 ~30 (≤80%R.H)

         30 ~40  (≤75%R.H)

         40 ~50 (≤45%R.H)

       Storage:

         -20 ~ +60  (≤80%R.H)

  Safety/ Compliances: 

    EN61326:1997+A1:1998+A2:2001+A3:2003, 

    EN61672-1: 2002 Class 2 and IEC60641:1979 Type 2, 

    ANSI S1.4:1983 Type 2

  Certification:   

Accuracy Specifications 

Accuracy:±(a% r b digits).guarantee for 1 year.

Operating temperature:23 ±5 .

Relative humidity:≤80%.

Temperature coefficient: 0.1 x (specified accuracy) / 1

A. Decibel dB

Maintenance

This section provides basic maintenance information 
including battery replacement instruction.

Warning

Do not attempt to repair or service your Meter unless 
you are qualified to do so and have the relevant 
calibration, performance test, and service information.  
To avoid electrical shock or damage to the Meter, do 
not get water inside the case.  

In order not to affect the Meter accuracy or damge to 
the Meter, do not open the Meter housing

 

A.General Service

  Periodically wipe the case with damp cloth and mild de

    tergent. Do not use chemical solvent.

  To clean the terminals with cotton-tipped swab with de

    tergent, as dirt or moisture in the terminals can affect 
    readings.

  Press the Meter power off when it is not in use and take 

    out the battery when not using for a long time.

  Do not store the Meter in place of humidity, high tem

    perature, explosive, inflammable and strong magnetic 
    field

B.Replacing the Battery 

(see figure 5)

 figure 5 

 

Warning

To avoid false readings replace the battery as 
soon as the battery indicator 

 appears.

To replace battery: 
1.Press the Meter power off
2.Remove the screw from the battery compartment, and 
   then take out the battery door from the battery compart
   ment.
3.Remove the battery from the battery compartment.
4.Replace the battery with 4pcs new 1.5V AA batteries
5.Rejoin the battery door and the battery compartment, 
   and install the screw

Copyright 2008 Uni-Trend Group Limited.

All rights reserved.

Manufacturer:

Uni-Trend Technology (Dongguan) Limited

Dong Fang Da Dao

Bei Shan Dong Fang Industrial Development District

Hu Men Town, Dongguan City

Guang Dong Province

China

Postal Code: 523 925

Headquarters:

Uni-Trend Group Limited

Rm901, 9/F, Nanyang Plaza

57 Hung To Road

Kwun Tong

Kowloon, Hong Kong

Tel: (852) 2950 9168

Fax: (852) 2950 9303

Email: [email protected]

http://www.uni-trend.com

14

MAX/

MIN

Function                         Range                             Resolution     Accuracy 

                  Remarks

Sampling Time: 125 microsecond

Sampling Time: 1 second

1 dB per scale, 

Sampling time: 200 times per second

Over range display: OVER

Under range display: UNDER

Has input terminal

Has input terminal

Turn on and off the Meter and data 

holding

Selecting auto and manual ranging

RECALL/CLR features are for model 

UT352 only.  Press to recall and clear 

data

STORE feature is for model UT352 

only.  Press to store data

30~80dB

50~100dB

60~110dB

80~130dB

FAST

SLOW

30~130dB

Output impedance around 

100Ω, 10mV/dB
Output impedance around 

600Ω, 0.707V/ each scale

A-Weighting and 
C-Weighting selection
Selection of fast or slow 

sampling rate and turn on 

and off of display backlight
Selection of maximum and 

minimum value.

Selection of auto power on 

and off

A-Weighting and 
C-Weighting

Sampling Rate

Analogue Bar Graph

Overloading 

DC analogue signal 
output
AC analogue signal 
output

Power (HOLD)

LEVEL (AUTO)

A/C (RECALL/CLR)

FAST/SLOW 
(STORE/BL)

MAX/MIN (Auto 
ranging)

0.1dB 

±1.5dB

 

 

 
 

1dB 

Frequency Response 

31.5~8kHz

    Press MAX/MIN button again.  The MIN 

      icon will appear on the display. The read

      ing displayed is the lowest reading

      encountered since the MIN mode was 

      entered.

    Press the MAX/MIN button again to exit 

      the MAX/MIN display mode

Отзывы: