B. Replacing the Battery 

(see figure 10)

Warning

To avoid false readings, which could lead to possible
electric shock or personal injury, replace the battery
as soon as the battery indicator “    ” appears.

Make sure the transformer jaw and the tets leads are
disconected from the circuit being tested before
opening the case bottom.

To replace the battery:

Turn the Meter off and remove all the connections from the input terminals
Turn the Meter’s front case down.
Remove the screw from the battery compartment, and separate the battery
compartment from the case bottom.
Take out the old battery and replace with a new 9V battery (NEDA1604, 6F22 or
006P.
Rejoin the case bottom and the battery compartment, and reinstall the screw.

(figure 10)

1.
2.
3.

4.

5.

** END **

This operating manual is subject to change without notice.

Screw

Model UT203/204: OPERATING MANUAL

45

Содержание UT203

Страница 1: ...Digital Clamp Multimeters P N 41416301 Registered Design No ZL02 3 357791 6 UT203 204 ...

Страница 2: ...Functional Buttons The Effectiveness of Functional Buttons Display Symbols Measurement Operation A DC AC Voltage Measurement B Measuring Resistance C Testing Diodes D Testing for Continuity E Frequency Measurement F Duty Cycle Measurement G DC AC Current Measurement 3 4 5 6 9 10 11 12 14 15 18 18 20 22 24 26 28 30 Title Page Model UT203 204 OPERATING MANUAL 1 ...

Страница 3: ...y Specifications A DC Voltage B AC Voltage C Resistance D Diode Test E Continuity Test F Frequency G Duty Cycle H DC Current I AC Current Maintenance A General Service B Replacing the Battery 32 33 33 34 35 35 36 37 37 38 39 40 40 42 44 44 45 Title Page Model UT203 204 OPERATING MANUAL 2 ...

Страница 4: ...Model UT203 UT204 hereafter referred to as the Meter are 3 3 4 digits with steady operations fashionable structure and highly reliable measuring instrument The Meter uses large scale of integrated circuit with double integrated A D converter as its core and has full range overload protection The Meter can not only measure AC DC Voltage AC DC Current Frequency Duty Cycle Resistance Diodes Continuit...

Страница 5: ...ee any missing or damaged part Item Description Qty 1 2 3 4 5 English Operating Manual Test Lead Test Clip Carrying Bag 9V Battery NEDA1604 6F22 or 006P 1 piece 1 pair 1 pair 1 piece 1 piece In the event you find any missing or damage please contact your dealer immediately Model UT203 204 OPERATING MANUAL 4 ...

Страница 6: ...smaller transient overvoltages than CAT IV Use the Meter only as specified in this operating manual otherwise the protection provided by the Meter may be impaired In this manual a Warning identifies conditions and actions that pose hazards to the user or may damage the Meter or the equipment under test A Note identifies the information that user should pay attention to International electrical sym...

Страница 7: ...ck the test leads for continuity Replace damaged test leads with identical model number or electrical specifications before using the Meter Do not apply more than the rated voltage as marked on the Meter between the terminals or between any terminal and grounding If the value to be measured is unknown use the maximum measurement position and reduce the range step by step until a satisfactory readi...

Страница 8: ...roper terminals function and range for your measurements Do not use or store the Meter in an environment of high temperature humidity explosive inflammable and strong magnetic field The performance of the Meter may deteriorate after dampened When using the test leads keep your fingers behind the finger guards Disconnect circuit power and discharge all high voltage capacitors before testing resista...

Страница 9: ...e of the Meter when servicing No abrasive and solvent should be used to prevent the surface of the Meter from corrosion damage and accident The Meter is suitable for indoor use Turn the Meter off when it is not in use and take out the battery when not using for a long time Constantly check the battery as it may leak when it has been using for some time replace the battery as soon as leaking appear...

Страница 10: ...ernating Current DC Direct Current AC or DC Grounding Double Insulated Warning Refer to the Operating Manual Deficiency of Built In Battery Continuity Test Diode Fuse Conforms to Standards of European Union Model UT203 204 OPERATING MANUAL 9 ...

Страница 11: ...the pressure on the lever is released the jaws will close Hand Guards to protect user s hand from touching the dangerous area Transformer Jaw designed to pick up the AC and DC current flowing through the conductor It could transfer current to voltage The tested conductor must vertically go through the Jaw center 1 2 3 4 5 6 7 Model UT203 204 OPERATING MANUAL 10 ...

Страница 12: ...positions Rotary Switch Position Power is turned off AC or DC voltage measurement Resistance measurement Diode test Continuity test Frequency Measurement and Duty Measurement AC and DC current measurement range Function OFF V Hz Duty 40A 400A Model UT203 204 OPERATING MANUAL 11 ...

Страница 13: ...Press to select manual ranging measurement mode The Meter is default to auto ranging measurement mode When the Meter is at manual ranging measurement mode press to step down the range At range Press to enter the REL mode It subtracts a stored value from the present measurement value and displays a result At Hz Duty range Press to switch between Hz measurement mode and Duty measurement mode l l l l...

Страница 14: ... the alternate functions marked in blue colour on the Meter s faceplate including Hz Duty 40 and 400 After the Meter entering Sleep Mode press and hold SELECT to turn the Meter on it will disable the Sleep Mode feature l l V A A Model UT203 204 OPERATING MANUAL 13 ...

Страница 15: ...d on every rotary switch positions Below table describe which functional buttons can be used on which rotary switch positions Rotary Switch Positions Functional Buttons SELECT REL HOLD l l l l l l l l l l l N A l l l l N A N A V Hz Duty 40A 400A Model UT203 204 OPERATING MANUAL 14 ...

Страница 16: ...Display Symbols see figure 2 figure 2 4 1 2 3 5 6 8 9 10 12 13 14 16 7 11 15 Model UT203 204 OPERATING MANUAL 15 ...

Страница 17: ...uld lead to possible electric shock or personal injury replace the battery as soon as the battery indicator appears The Meter is in the auto range mode in which the Meter automatically selects the range with the best resolution Test of diode The continuity buzzer is on Indicator for Duty Data hold is active Indicator for REL mode Model UT203 204 OPERATING MANUAL 16 ...

Страница 18: ...103 or 1000 ohms M Megohm 1x106 or 1 000 000 ohms The unit of Frequency Amperes amps The unit of current Volts The unit of voltage mV Millivolt 1x10 3 or 0 001 volts Indicates negative reading Indicator for TRMS mode The input value is too large for the selected range Model UT203 204 OPERATING MANUAL 17 ...

Страница 19: ...asure voltages higher than 600V AC DC although readings may be obtained The DC Voltage ranges are 400mV 4V 40V 400V and 600V The AC Voltage ranges are 4V 40V 400V and 600V To measure DC voltage connect the Meter as follows Insert the red test lead into the Hz Duty terminal and the black test lead into the COM terminal 1 Red Black Model UT203 204 OPERATING MANUAL 18 ...

Страница 20: ... from the input terminals Set the rotary switch to DC mesaurement mode and auto ranging is a default Press SELECT to switch to AC measurement mode or press to switch to manual ranging measurement mode Connect the test leads across with the object being measured The measured value shows on the display 2 3 V REL Model UT203 204 OPERATING MANUAL 19 Note l ...

Страница 21: ...400 4k 40k 400k M and 40M To measure resistance connect the Meter as follows Insert the red test lead into the Hz Duty terminal and the black test lead into the COM terminal 1 2 3 Set the rotary switch to Resistance measurement is default to auto range mode press to switch to manual ranging measurement mode Connect the test leads across with the object being measured The measured value shows on th...

Страница 22: ...eing tested from the circuit when measuring When resistance measurement has been completed disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals Model UT203 204 OPERATING MANUAL 21 Note l l ...

Страница 23: ...ransistors and other semiconductor devices The diode test sends a current through the semicondutor junction then measure the voltage drop across the junction A good silicon junction drops between 0 5V and 0 8V To test the diode out of a circuit connect the Meter as follows Insert the red test lead into the Hz Duty terminal and the black test lead into the COM terminal 1 figure 5 Red Black Model UT...

Страница 24: ...nd the circuit under test and remove testing leads from the input terminals Set the rotary switch to Diode measurement mode is a default or presss SELECT to select measurement mode For forward voltage drop readings on any semiconductor component place the red test lead on the component s anode and place the black test lead on the component s cathode 2 3 Model UT203 204 OPERATING MANUAL 23 Note l l...

Страница 25: ...Insert the red test lead into the Hz Duty terminal and the black test lead into the COM terminal Set the rotary switch to and press SELECT button to select measurement mode The buzzer sounds if the resistance of a circuit under test is less than 50 1 2 3 4 5 The buzzer may or may not sounds if the resistance of a circuit under test is between 50 to 100 The buzzer does not sound if the resistance o...

Страница 26: ...ontinuity testing has been completed disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals Model UT203 204 OPERATING MANUAL 25 Note l ...

Страница 27: ...tance ranges are 10Hz 100Hz 1kHz 10kHz 100kHz 1MHz and 10MHz To measure frequency connect the Meter as follows Insert the red temperature probe into the Hz Duty terminal and the black temperature probe into the COM terminal 1 2 3 figure 7 Red Black Set the rotary switch to Hz Connect the test leads across with the object being measured The measured value shows on the display Model UT203 204 OPERAT...

Страница 28: ...urement has been completed disconnect the connection between the testing leads and the circuit under test and remove the testing leads away from the input terminals of the Meter Model UT203 204 OPERATING MANUAL 27 Note l ...

Страница 29: ...d The duty cycle range is 0 1 99 9 To measure duty cycle connect the Meter as follows Insert the red temperature probe into the Hz Duty terminal and the black temperature probe into the COM terminal 1 2 3 REL Set the rotary switch to Hz and press to select Duty Cycle measurement mode Connect the test leads across with the object being measured The measured value shows on the display Model UT203 20...

Страница 30: ...surement has been completed disconnect the connection between the testing leads and the circuit under test and remove the testing leads away from the input terminals of the Meter Model UT203 204 OPERATING MANUAL 29 Note l ...

Страница 31: ...net but also to heat and machines reaction force Any shock will cause the changing in reading in the short time Press the lever to open the transformer jaw A A A A Center the conductor within the transformer jaw then release the Meter slowly until the trasnformer jaw is completely closed Make sure the conductor to be tested is placed at the center of the transformer jaw otherwise it will casue dev...

Страница 32: ...nt value and displays a result When current measurement has been completed disconnect the connection between the conductor under test and the jaw and remove the conductor away from the transformer jaw of the Meter REL Model UT203 204 OPERATING MANUAL 31 Note l l ...

Страница 33: ...pressing the button based on The Effectiveness of Functional Buttons on page 14 Press SELECT to activate the Meter will disable the Sleep Mode feature The Meter beeps 5 times in about 1 minute before entering Sleep Mode and it will have a 1 long beep just before entering Sleep Mode To disable the Sleep Mode function press and hold SELECT button while turning on the Meter Model UT203 204 OPERATING ...

Страница 34: ...econd Measuremnet Deviation The conductor being meaured is not placed in the center of the jawt during AC DC current measurement it will cause extra 1 deviation based on the stated accuracy Drop Test 1 meter drop test passed Max Jaw Size 28mm diameter Projected Max Current conductor size 26mm diameter Electro Magnetic When carrying out measurement near the electro magnetic it may cause unstable or...

Страница 35: ... l l B Environmental Requirements The Meter is suitable for indoor use Altitude Operating 2000m Storage 10000m Safety Compliances IEC 61010 CAT II 600V CAT III 300V over voltage and double insulation standard Temperature and humidity Operating 0ºC 30ºC 85 R H 30ºC 40ºC 75 R H 40ºC 50ºC 45 R H Storage 20ºC 60ºC 85 R H l l l l Model UT203 204 OPERATING MANUAL 34 ...

Страница 36: ...e 23ºC 5ºC Relative humidity 85 R H Temperature coefficient 0 1x specified accuracy 1ºC A DC Voltage Range Resolution Accuracy Overload protection 400 0mV 4 000V 40 00V 400 0V 600V 0 1mV 1mV 10mV 100mV 1V 0 8 3 0 8 1 1 3 600V DC AC Remarks Input impedance 10M Model UT203 204 OPERATING MANUAL 35 ...

Страница 37: ... by using average response method Input sine wave then adjust the reading until it is same as the effective value Combine AC and True RMS response method Input sine wave to adjust Non sine wave must follow the below data to adjust Peak factor 1 4 2 0 add 1 0 on the stated accuracy Peak factor 2 0 2 5 add 2 5 on the stated accuracy Peak factor 2 5 3 0 add 4 0 on the stated accuracy UT203 UT203 Mode...

Страница 38: ...k 40 00k 400 0k 4 000M 40 00M 100m 1 10 100 1k 10k 1 2 2 1 2 1 2 2 1 5 2 600Vp D Diode Test Range Resolution Accuracy Overload protection 1mV Display forward voltage drop nearest value 600Vp Remark Open circuit voltage approximate 1 48V Model UT203 204 OPERATING MANUAL 37 ...

Страница 39: ...the buzzer beeps 600Vp Remark Open circuit voltage approximate 0 45V The buzzer may or may not beeps when the resistance of a circuit under test is between 50 100 The buzzer will not beep when the resistance of a circuit under test is 100 l l l Model UT203 204 OPERATING MANUAL 38 ...

Страница 40: ...ction 10Hz 100Hz 1kHz 10kHz 100kHz 1MHz 10MHz 0 001Hz 0 01Hz 0 1Hz 1Hz 10Hz 100Hz 1kHz 0 1 3 600Vp For reference only Remark Input Sensitivity as follows When 100kHz 300mV rms When 100kHz 600mV rms When 1MHz 800mV rms Model UT203 204 OPERATING MANUAL 39 ...

Страница 41: ...0 00A 400 0A 2 5 2 3 Warning The operating temperature must be 0ºC 40ºC when measuring current Remark If the reading is positive the current direction is from bottom to up See figure 10 the front case face up while the bottom case face down Hold the Meter tight do now release The Hall components are very sensitive not only to the magnet but l Model UT203 204 OPERATING MANUAL 40 ...

Страница 42: ...nter the conductor within the transformer jaws then release the Meter slowly until the transformer jaw is completely closed Make sure the conductor to be tested is placed at the center of the transformer jaw otherwise it will cause 1 0 deviation based on the stated accuracy Remove the transformer jaw Press to display zero Repeat the above 1 procedure The obtained reading will be more precise l l l...

Страница 43: ... not only to the magnet but also to heat and machines reaction force Any shock will cause the changing in reading in the short time Follow the below procedure to measure current will be more precise l l Hold the Meter tight and press the lever to open the transformer jaw Center the conductor within the transformer jaws then release the Meter slowly until the transformer jaw is completely closed Ma...

Страница 44: ...to adjust Non sine wave must follow the below data to adjust Peak factor 1 4 2 0 add 1 0 on the stated accuracy Peak factor 2 0 2 5 add 2 5 on the stated accuracy Peak factor 2 5 3 0 add 4 0 on the stated accuracy Remove the transformer jaw Press to display zero Repeat the above 1 procedure The obtained reading will be more precise 2 3 4 5 l Change to AC REL Model UT203 204 OPERATING MANUAL 43 ...

Страница 45: ...not get water inside the case A General Service Periodically wipe the case with a damp cloth and mild detergent Do not use abrasives or solvents To clean the terminals with cotton bar with detergent as dirt or moisture in the terminals can affect readings Turn the Meter power off when it is not in use Take out the battery when it is not using for a long time Do not use or store the Meter in a plac...

Страница 46: ...tom To replace the battery Turn the Meter off and remove all the connections from the input terminals Turn the Meter s front case down Remove the screw from the battery compartment and separate the battery compartment from the case bottom Take out the old battery and replace with a new 9V battery NEDA1604 6F22 or 006P Rejoin the case bottom and the battery compartment and reinstall the screw figur...

Страница 47: ...Dong Fang Industrial Development District Hu Men Town Dongguan City Guang Dong Province China Postal Code 523 925 Headquarters Uni Trend Group Limited Rm901 9 F Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon Hong Kong Tel 852 2950 9168 Fax 852 2950 9303 Email info uni trend com http www uni trend com Model UT203 204 OPERATING MANUAL 46 ...

Отзывы: