I. Safety Information

     In order to correctly use this instrument, please read this manual carefully  

in detail. This manual includes the safety information related to the safe oper

-ation  of the Tester, please strictly observe them when using the Tester.

 

Caution:

1. Please read and understand the contents of this manual before use.

2. Please keep this manual properly. It can be easily used for a test reference 

    when necessary. 

    
3. Ensure to use the  instrument in accordance with the test procedures  

    described in this manual.
4. Make sure you have understood the safety information details.

5. Ensure to strictly follow the safety regulations, or otherwise accidents or   

    instrument damage may occur. 

 

 

6. The Instrument should be operated by qualified and trained personnel and  

     used only as specified in the manual. 

     

 

 

 

7.  UNI-T will not be liable for any damage to the instrument or losses caused by

misuse or operation not as specified in the manual or the safety regulations.

 

 

 

 

   

 

    

Safety mark      has three meanings in  the manual; when reading, user  
will pay special attention to the operation relating to      .

Danger: Specifies conditions and actions that most likely pose 

hazards to the user.

Caution: Specifies conditions and actions that may cause minor 

injury or damage to the Tester

Warning: Specifies conditions and actions that may pose hazards 

to the user;

This instrument is only for the single phase within AC 230V/50Hz

195~253V

.

When testing, do not touch any bared wire.

When testing, you can touch the tested conductor only in the safe  

situation.

After the test is finished, Please disconnect the test lead from the 

power supply. Do not leave the test lead connected to the power 

supply for a long time.

Danger

Warning

Do not perform the tests with the Tester  cover opened, for hazardous 

voltage may be present. If the Tester work abnormally, please have it 

checked and serviced by professional personnel.

If any anomaly happen(eg: missing display, unwanted data, casing 

damage, test noise, etc), please contact the qualified servicing 

personnel before proceeding any operation.

 

Before using the instrument, if your hand is wet, please do not use it.

please use only the certified test lead supplied with the Tester, prohibit 

using other kinds of test wires.

Do not expose the instrument under sunshine, extreme temperature,  

moisture and other severe environment.

Clean the Tester with dampened cloth and mild detergent; no abrasives 

or solvents are allowed.  

 

When the instrument is wet, please ensure it  is dried before storage.

Caution

Electrical Symbols

Marks of danger, warning and caution

Double or reinforced insulation

CE conforms to standard of European Union.

II. Product Features

2.0

 Adopt intellectual micro-controller chip to maintain high accuracy,  

 high reliability and stability.  

2.1  Wiring inspection (RCD test ):

2.2

 

Phase angle selection:

The tests can be performed at positive 0  or negative 180   phase angle.

2.3

 

Test locked

 

When the test lead is disconnected, long press 0 / 180 (

  )

button to enable 

the test function, then connect the test lead to tested terminals, the tests can 

be automatically performed.

2.4

 

AUTO RAMP test: 

Measure trip current and trip time simultaneously.

2.5

 

Over-Range Indication: 

If the Tester fails to trip within Max. test time, the display shows     

 

icon.

2.6

 

Data Hold:

The readings will be frozen when the test finishes.

2.7

 

Maximum current of test is up to

 

1A ( x1 Function).

2.8

 

Over-heat inspection: 

When the related parts inside the instrument are over heated, LCD  displays 

and the instrument will automatically prohibit the tests, after it is cooled 

 

down, LCD does not display 

and  

  

the tests restart again.

2.9

 

Backlight : 

Press LIGHT(    ) to turn on /off the backlight. When under MANUAL RAMP 

mode, press the button momentarily to turn on/off the backlight, or long press 

to access I

n setup.

 

2.10

 

Low battery indication:

When voltage of battery is less than 7V, low-voltage sign 

     

is displayed,  

please change the battery in time;

2.11

 

Auto power off : 

The instrument will automatically power off after ten minutes' idling. To disable 

the function, press and hold 0  / 180  button while turning on the instrument.

2.12

 

Fused protection.

III. Technical Specifications

3.1

 

Measurement Range and Accuracy:

Temperature: 23 5
Humidity:: 45%~75%RH

Altitude:   2,000m

Accuracy: ±(a% of reb digits) 

10/20/30/100/200

/300/500mA/1000mA

10/20/30/100/200

/300mA

500mA/1000mA

0~2000mS 0 /180

0~2000mS 0 /180
0~200mS  0 /180
0~200mS  0 /180

10/20/30/100/200

/300mA/500mA

10/20/30/100/200

/300/500/1000mA

10/20/30/100/200

/300/500/1000mA

0

/

180

0~200mS

10/20/30/100/200mA

-10%~0%

0.6% 

rdg 

4dgt

-10%~+10%

-10%~+10%

0%~+10%

195~253V

/50Hz

Auto 

Ramp

Manual 

Ramp

Function

Rated Trip Current

(I

n) 

Working 

voltage (AC)

Accuracy

Trip Time

Trip Time

Trip Current

1V

60 400V 

500VAC/700VDC

5%+2

Function

 Range

Resolution Accuracy

Input 

Protection

Remarks

AC 

voltage

Input Impedance 

about 5.9M

LCD Indication                                                  Connection Status

40%~110% of Rated 

Current, step at 5%; 

300ms ×5;0  / 180 

             

L-N&L-PE flash                                  Abnormal power supply or no supply to 

                                                           tested circuits

L-PE&N-PE flash, L-N doesn’t           The tested terminals are badly or not 

                                                           earthed.

L-N&N-PE flash, L-PE doesn’t           Neutral terminal is disconnected

L-N,L-PE and N-PE flash                   Wrong wiring to tested terminals

The tester will prohibit the tests regarding the above status, except reverse 

connection between earth and neutral terminals,   