![UNI-T UPO3000E Series Скачать руководство пользователя страница 54](http://html2.mh-extra.com/html/uni-t/upo3000e-series/upo3000e-series_programming-manual_3712431054.webp)
UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:TRIGger:SHOLd:TIME:STEP
Command format:
:TRIGger:SHOLd:TIME:STEP <step>
:TRIGger:SHOLd:TIME:STEP?
Functional description:
Set time interval of setup hold trigger to adjust stepped value.
Return format:
Query return the current stepped value, unit is s.
For example:
:TRIGger:SHOLd:TIME:STEP 0.001
set time interval of setup hold trigger as 1ms
:TRIGger:SHOLd:TIME:STEP?
query return 1.000e-003
:TRIGger:SHOLd:TIME
Command format:
:TRIGger:SHOLd:TIME <time>
:TRIGger:SHOLd:TIME?
Functional description:
Set the time interval of setup hold trigger.
Return format:
Query return the current time interval, unit is s.
For example:
:TRIGger:SHOLd:TIME 1
set time interval of setup hold trigger as 1ms
:TRIGger:SHOLd:TIME?
query return 1.000e000
:TRIGger:SHOLd:SELect
Command format:
:TRIGger:SHOLd:SELect
<SOURce<n>>
:TRIGger:SHOLd:SELect
Functional description:
For switch selected source, SOURce<n> presents source, n take value as1, 2.
SOURce1 presents data source
;
SOURce2 presents clock source.
Return format:
Query return{ SOURce1 | SOURce2 }.
For example:
:TRIGger:SHOLd:SELect SOURce1
set the selected focus source
:TRIGger:SHOLd:SELect?
query return SOURce1