You can use three channels of 0~10V analog signals to control the output on/off of the voltage, current and 
power respectively; Vm and Im channels output 0~10V analog signals to monitor the voltage and current 
output of the power supply respectively. 
Terminal explanation as show in in the following table.   

 

 

Connecting Example of External Analog Control Hardware 

 

(1)

 

As shown in the above figure, entering the output settings to set the operating mode to the 
external analog control Ext-V. 

(2)

 

Slide to Rv or RI to set the value of voltage or current. 

(3)

 

Close S1 to turn on the output.   
 

3.15.4 RS232 and RS485 

RS232 DB9 and RS485 are using the same serial port. 
 

 

Normal Mode 

External Analog Control Ext-V 

External Digital Control 

Ext-D 

Ref 

10V Regulated Voltage Output 

Vs/D0 

Null 

Voltage analog control (0~10V) 

Digital input D0 

Is/D1 

Null 

Current    analog control 

(0~10V) 

Digital input    D1 

D2 

Null 

Null 

Digital input t D2 

ON 

Null 

Power output the external control, high: turn on the 

output; low: turn off the output 

Vm 

Voltage monitor analog quantity output (0~10V) 

Im 

Current monitor analog quantity output

(0~10V) 

Содержание UDP6900 Series

Страница 1: ...User s Manual UDP6900 Series Digital Control Power Supply...

Страница 2: ...ne year from the date of purchase If the instrument is damaged due to improper operation by the user during the warranty period the maintenance fee and the costs caused by the maintenance shall be bor...

Страница 3: ...nsufficient maintenance UNI T is not obliged to provide the services below as prescribed by the warranty a Repair damage caused by installation repair or maintenance of personnel other than service re...

Страница 4: ...ot open the cover during operating Do not operate the instrument when open the instrument cover and do not change the internal circuit Do not contact live part When the instrument is operating do not...

Страница 5: ...perating Environment Requirements Operating temperature 0 40 Operating humidity 20 80 non condensation Storage temperature 10 70 Altitude 2000 meters Pollution degree 2 Cleaning Please pull out power...

Страница 6: ...r distributor Article Quantity Remarks UDP6900 series DC programming power supply 1 piece The model is subject to the actual order 3C power cable 1 piece Calibration report 1 piece Calibration report...

Страница 7: ...ni trend com 7 44 1 3 Handle Handle can adjust to three positions by appropriate strengths Hold the handle and pull to two sides to remove it Adjusting the handle to the position as shown in the follo...

Страница 8: ...saves your cost and space Take UDP6942B 60V 15A 360W as an example it can realize various combination of 60V 6A 360W 40V 9A 360W and 24V 15A 360W UDP6900 series has a built in interface of RS232 RS48...

Страница 9: ...XI bus Supports 100 to 240V input voltage Model Voltage Current Power UDP6922B 60V 5A 100W UDP6932B 60V 10A 200W UDP6933B 150V 5A 200W UDP6942B 60V 15A 360W UDP6952B 60V 25A 600W UDP6953B 150V 10A 600...

Страница 10: ...Long press to save the current page to USB memory when USB is connected List Enable and switch list output or delay function Lock Short press to lock long press to unlock the keyboard Preset Save the...

Страница 11: ...twork cable USB is plug in and recognized Display the output voltage Display the output current Display the reading when the voltmeter function has been used Normal The power operating mode is normal...

Страница 12: ...upply please make sure that you have fully understood the safety instructions Please make sure the power voltage meets the power supply otherwise the power supply may be burned out Please make sure th...

Страница 13: ...power switch is switched on then move to next step If the power supply is turned off press the power switch to fix the error 3 Check whether the fuse is burned out If the fuse is burned out replace th...

Страница 14: ...gs The voltage range is between 0V to the maximum output When press the voltage key the cursor indicator will move to the voltage field as shown in the following figure There are two ways to adjust th...

Страница 15: ...ge settings 3 4 Output ON OFF Press the output key to turn on the function and the indicator is green The user s page will highlight to display the output voltage current ON OFF state power and output...

Страница 16: ...CV mode and is indicated by the CV status text If the output current is limited by the current settings value or rated current the product switches to constant current mode CC and is indicated by the...

Страница 17: ...column use the arrow key with the encoder knob to switch to the set value or use numeric keyboard to input the set value push the encoder knob one time the icon represents the voltage protection valu...

Страница 18: ...the specified file from the internal Flash storage or external USB storage The storage format is lst csv and dly csv User can directly open and quickly edit the EXCEL file on computer 3 11 1 List Out...

Страница 19: ...Edit button to set the maximum and minimum of voltage and current high low level pulse width period phase of the waveform 5 After the above settings is finished click on the Generate button to complet...

Страница 20: ...is decided by the high level and the low level 3 Ramp Wave The difference between the maximum and minimum values and the time interval maximum 9999 9 determines the slope of the ramp wave The symmetr...

Страница 21: ...e step falling is N 1 2 1 when N is even number the step rising is N 2 the step falling is N 2 The time interval maximum 9999 9 represents the duration of each step The time interval and the total cou...

Страница 22: ...d as x is the independent variable and can be taken as an integer from 0 to P 1 3 11 2 Delayer Settings 1 Press the list key on the panel multiple times to enter Delayer page 2 Click on the basic sett...

Страница 23: ...elayer of this time the maximum sum of the start point and the total point is 512 4 State code is used to set the template state 5 Duration template is used to set the time change Fixed duration repre...

Страница 24: ...hat extends the control output channel to the open or close state by increasing the step value Tstep 9999 9s based on Tbase Calculation formula T i 1 T i Tstep T 0 Tbase PS This template only changes...

Страница 25: ...t Note the file suffix is lst csv After enter the delayer function click on the Manage button to save or read the delayer data Storage Interface of Delayer Note the file suffix is dly csv Press the en...

Страница 26: ...tton to complete reading and using the stored files Read Interface of List Output File 3 12 Main Menu The main menu includes the following contents Monitor Output Settings System Settings User Group L...

Страница 27: ...ngs whether to turn off the output whether to pop up the prompt box and whether to turn on the beeper 4 Turn on the monitor function Click on the Start button to turn on the function 3 12 2 Output Set...

Страница 28: ...current is set by manual or remote command Externalanalogcontrol Ext V thevoltage and current settingvalues change accordingtothevoltage values Vs and Is input from the rear interface External digital...

Страница 29: ...backlight brightness setting after the setting is finished Note The range of backlight brightness is 20 100 Beeper Setting Enter the basic setting press the button slightly under Sound to locate to t...

Страница 30: ...e encoder knob to switch and select a protocol 4 The communication address can set to 1 32 Address 0 is used to broadcast only valid in the protocol of Multi SCPI and Modbus 5 Click on the ESC key to...

Страница 31: ...r knob to select a group press the button under Save and then press the button under Load to modify the name Press the button under Save then press the button under Load and press the Rename button to...

Страница 32: ...er output GND Ref 10V regulated power supply output it is generally used for external analog control Ext V Vs D0 External analog control Voltage analog quantity 0 10V is corresponding to 0 VoutMax Is...

Страница 33: ...The maximum measurement voltage of UDP6900 series is 65V the maximum measurement voltage of UDP6933B 53B is 160 0V If a voltage is high thanthe limit value it may cause damage to the instrument 3 Volt...

Страница 34: ...y it can control the output voltage current and switching output by inputting an external voltage 0 10 V and output monitor by outputting 0 10 V through Vm and Im If the user connects a voltage contro...

Страница 35: ...the output settings to set the operating mode to the external analog control Ext V 2 Slide to Rv or RI to set the value of voltage or current 3 Close S1 to turn on the output 3 15 4 RS232 and RS485 R...

Страница 36: ...d data bits no parity check 8 bits of data bits 1 bit of stop bit Communication Settings 1 Enter the menu settings 2 System Settings Communication Settings adjust the baud rate communication protocol...

Страница 37: ...Manual UDP6900 Series Digital Control Power Supply Instruments uni trend com 37 44 Note USB to RS485 module is required to connect to a PC 4 Remote Control 4 1 Upper Computer Control Remote Operation...

Страница 38: ...elect the connection method andthendouble clicktheUDP6942Bpower supply icontoentertheremote communication interface The operation diagram is as follows 1 Click on the Scan button to pop out the select...

Страница 39: ...ly to enter System Settings and Network Settings Note When DHCP is turned on it will automatically obtain IP address and update IP address and other parameters requires the router to support DHCP serv...

Страница 40: ...er defined parameters Submit the settings and refresh the setting parameters Turn on off List function Set and submit the group parameter settings Page up down refresh group parameter and submit group...

Страница 41: ...User s Manual UDP6900 Series Digital Control Power Supply Instruments uni trend com 41 44...

Страница 42: ...10mV 0 01 30mV Current 0 05 2mA 0 05 4mA 0 05 6mA Setting Resolution Voltage 1mV 1mV 1mV Current 0 1mA 0 1mA 0 1mA Readback Resolution Voltage 0 1mV 10V 1mV 10V 0 1mV 10V 1mV 10V 0 1mV 10V 1mV 10V Cur...

Страница 43: ...ffset Voltage 0 04 30mV Current 0 1 10mA Readback accuracy 25 C 5 C of output offset Voltage 0 04 30mV Current 0 1 10mA Ripple Voltage 30mVp p Current 6mArms SENSE voltage V 1V Size 215mmW 88mH 373 7m...

Страница 44: ...mA Readback accuracy 25 C 5 C of output offset Voltage 0 03 5mV 0 03 20mV Current 0 1 25mA 0 1 25mA Ripple 20Hz 20MHz Voltage 20mVp p 50mVp p Current 15mArms 15mArms SENSE voltage V 1V Size 215mmW 88m...

Отзывы: