Quick Start Guide MSO/UPO3000E Series Oscilloscope
Instruments.uni-trend.com
8 / 17
contact UNI-T or local distributors of this product.
(3) Machine Inspection
If the instrument appears to be damaged, not working properly, or has failed the functionality test, please
contact UNI-T or local distributors of this product.
If the equipment is damaged due to shipping, please keep the packaging and notify both the transportation
department and UNI-T distributors, UNI-T will arrange maintenance or replacement.
3.2 Before Use
To perform a quick verification of the instrument's normal operations, please follow the steps below:
(1) Connect to the Power Supply
The power supply voltage range is from 100 VAC to 240 VAC, the frequency range is 50 Hz to 60 Hz. Connect
the oscilloscope to the power supply line that came with the oscilloscope or any power supply line that meets
the local country standards. If the power switch on the rear panel is not turned on, the indicator of power soft
button
on the left corner on the front panel will be extinguished and no function. If the power switch
on the rear panel is turned on, the indicator of power soft button
on the left corner on the front panel
will be on red, press this button to boot-up the instrument.
(2) Boot-up Check
Press the soft power button
and the light should change to green. The oscilloscope will show a boot
animation, and then enter the normal interface.
(3) Connect Probe
Use probe’s BNC port connect to the channel 1 BNC port of the oscilloscope. Connect the probe to the
“Compensating signal port” and the ground clip is connected to the “Ground terminal” below the
compensating signal port as shown in Figure 1. The output of the compensating signal should be amplitude
3 Vpp, default frequency is 1 kHz
Figure 1 Compensating signal and Ground terminal
(4) Function Check
Press the AUTO key, a 3 Vpp 1 kHz square wave should appear. Repeat step 3 to check all channels.
(5) Probe Compensation
When the probe is connected to any input channel for the first time, this step might be adjusted in order to
Compensating signal terminal
Ground terminal