Function 50:
Saving the system configuration (system.cfg) on the SD card. Necessary
before a
firmware update.
Function 51:
Write back of the on SD card stored system configuration (system.cfg). Necessary
after a
firmware update.
Function 52:
This function allows the existing hardware and software version and thus the revision
level are displayed.
Function 53:
This function is used to generate a user-defined password to protect a
unauthorized entry of parameters.
That is, when selecting the parameters input by "B" will prompt you to
Entering a four-digit password (in the delivery of 0000).
In the parameter input is now on the function: 53 the creation of a
customer password possible.
Here is a unique password generated, so this password is in the future
Selection of the parameters input by "B" valid.
Function 66:
In this function, the values of the factory generated calibration value of the
analog current and / or voltage outputs can be displayed.
Function 80:
Reset of Totalizer (select Totalizer or Totalizer per day).
Function 90:
Automatic zeroing V and H (currently at the V and H – input pending value
is called as zero).
Function 91:
Automatic zeroing V (currently on the V - input pending value is set as zero).
Function 92:
Automatic zeroing H (currently on the H - input pending value is set as zero).
All three functions (90, 91 and 92) set for the selected parameter the currently measured input value as
zero (for display and calculation).
Function 93:
Zeroing to 4mA, zero for V and H are set exactly 4mA.
Function 94:
Zeroing to 4mA, zero for H are set exactly 4mA.
Function 95:
Zeroing to 4mA, zero for V are set exactly 4mA.
All three functions (93, 94 and 95) for the selected parameter set 4 mA as zero (for display and
calculation).
This function can also be used to correct an erroneous zero point adjustment function with 90,91 or 92.
Function 98:
Correction of the V - curve (depending on the filling level). Height-dependent V-correction
in 5 steps with entering the height cutoff point and the addition in %.
Function 99:
Displaying Date and Time