Initial operation
Initial operation | Setting zero point
| 3
Put the KSAT VIEW CD into your
computer or download software
from
www.ums-muc.de/KSAT.zip
.
Double click
ksat.msi
and follow
the installation wizard.
Connect the water supply and
discharge hoses.
Connect the device to your computer's
USB port.
The wizard assists you through
the installion.
You are ready to measure!
The device connects automatically
with your computer.
Start the KSAT VIEW software.
Note
For installing the KSAT VIEW software you may need administrator rights.
Fill burette by opening the fill cock, then
close it.
Fill measuring dome by opening the
burette cock.
Put wiper plate on the water lense and
and take it horizontally off.
Select „Setting zero point“ in the software.
Select „Setting zero point“ button.
A water lense is visible.
Water level is at the rim
of the measuring dome.
Setting zero point
Setting
zero point
Note: pressure reading
In the mode „measuring“ the screen shows the value -6.9 cm (-3 in) water column after
setting zero point. This is because the measuring setup is 6.9 cm (3 in) high.