28
4
Calibrating the capacity point
Load the calibration weight in the center
of the balance.
The display flashes, indicating that
capacity point calibration is in progress.
5
Terminating the span adjustment
When the calibration of the capacity
point is finished, the original
measurement mode is restored.
☆
Key Points of the Procedure
☆
1.
Pressing the Function key in Step 2 interrupts the span adjustment and returns you to the
original measurement mode.
2.
The calibration weight used for span adjustment should be heavier than 50% of the capacity of
the balance.
To implement a calibration as precisely as possible, use a weight close to the capacity of the
balance.
※
Calibration weights can be purchased from Shinko Denshi or local dealers.
For detail information, please contact Shinko Denshi or local dealers.
3.
If problems arise during span adjustments, one or more of the following error messages will
appear:
(1) [o-Err]:
The calibration weight exceeds the capacity of the balance.
(2) [1-Err]:
The calibration weight is less than 50% the capacity of the balance.
(3) [2-Err]:
The difference between before and after calibration values is too large (1.0%
or more).
※
If error messages are displayed, calibration cannot take place.
Check the weight and re-calibrate. If the same error continues after repeated calibrations using
the correct weight, please contact our Technical Service Division or your local dealer.