Calibrating the capacity point 

 

 

Load the calibration weight in the center 
of the balance. 
The display flashes, indicating that 
capacity point calibration is in progress. 

Terminating the span adjustment 

 

 

When the calibration of the capacity 
point is finished, the original 
measurement mode is restored. 

  Key Points of the Procedure 

 

1. 

Pressing the    Function    key in Step 2 interrupts the span adjustment and returns you to the 
original measurement mode. 

2. 

The calibration weight used for span adjustment should be heavier than 50% of the capacity of 
the balance.   
To implement a calibration as precisely as possible, use a weight close to the capacity of the 
balance.   

  Calibration weights can be purchased from Shinko Denshi or local dealers. 

For detail information, please contact Shinko Denshi or local dealers. 

3. 

If problems arise during span adjustments, one or more of the following error messages will 
appear: 

(1)  [o-Err]: 

The calibration weight exceeds the capacity of the balance. 

(2)  [1-Err]: 

The calibration weight is less than 50% the capacity of the balance.   

(3)  [2-Err]: 

The difference between before and after calibration values is too large (1.0% 
or more).   

If error messages are displayed, calibration cannot take place. 

Check the weight and re-calibrate.    If the same error continues after repeated calibrations using         
the correct weight, please contact our Technical Service Division or your local dealer. 

Содержание HYPROP

Страница 1: ...lectronic Balance HYPROP BALANCE Operation Manual To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe place near the balance so...

Страница 2: ...33 UMS GmbH 2014...

Страница 3: ...perate and features user friendly keys Furthermore the large liquid crystal display provides excellent visibility and the instrument s high speed and stability intrinsic to a tuning fork design help b...

Страница 4: ...imit Function Setup 22 8 2 Setup of Limit Values by Actual Quantity Loads 24 8 3 Setting up Limit Values by Inputting Values 25 9 Calibrating the Balance 9 1 Calibration of HYPROP BALANCE Series 27 9...

Страница 5: ...ENDED This term indicates steps that the user should take to ensure the quality and reliability of the balance Meanings of Symbols Each symbol is accompanied by an instruction Mandatory Symbol Indicat...

Страница 6: ...ric shock Keep Dry Do not use the balance in a location were it may be subjected to excess moisture Electric shock or short circuiting could occur The balance may be corroded with resultant malfunctio...

Страница 7: ...re increase in the balance may lead to inaccurate measurement Unplug Adapter If the balance is to be unused for an extended period of time unplug the adapter This conserves power and prevents deterior...

Страница 8: ...lance Turn the adjusters until the bubble rests in the center of the red circle Adjusters Total of four one on either side at front and rear AC adapter connector Measurement pan Liquid crystal display...

Страница 9: ...ues from memory If a value is flashing it is being saved CAL Stays on and flashes while span adjustment is in progress Auxiliary scale interval Lights up only when the auxiliary scale interval is disp...

Страница 10: ...mit function Function key Brief press toggle switches the units to be displayed in succession g Pcs etc Brief press moves the flashing digit in the setup of a limit value when using the value input me...

Страница 11: ...il the bubble rests in the center of the red circle on the level Caution Use caution when operating the adjusters on the square pan type to prevent them from lifting up 3 Connecting the AC adapter Con...

Страница 12: ...witch is set on After every segment lights If something is loaded the display does not change on 0 In this case remove the load 3 Switching the measurement mode Press the Function key Each time the ke...

Страница 13: ...dicate only the weight of the sample loaded in the tare Weighing only the weight of an added sample 4 Resetting the indication to zero Press the Zero Tare key The total weight of the sample loaded on...

Страница 14: ...r other vibrations Use the windshield or vibration dampers to protect against such adverse effects 4 When the read out indicator is reset to zero or the tare is subtracted the balance indicates zero t...

Страница 15: ...the Function key The indication changes to the next item 2 SEL 0 Limit function 3 Selecting an item Pressing the Function key advances the function items at the rate of one item per press 4 Changing...

Страница 16: ...HI OK and LO Auto zero zero tracking 3 A 0 0 Disable This function automatically sets the zero point exactly to zero to prevent slight deviations 1 Enable Auto power off 4 A P 0 Disable balance opera...

Страница 17: ...ANCE Series 4 3 Interface Section Displayed when 7 I F is set to 1 or 2 Item Set Value Description Output Control 71 o c 0 Stop output 1 Output continuous at all times 2 Output continuous if stable st...

Страница 18: ...of Measurement Only when the lock switch is off When values 81 S u to 85 S u are entered prior to use the desired unit of measurement to be displayed can be chosen simply by pressing the Function key...

Страница 19: ...n be set please refer to 4 2 Description of Functions on page 13 2 The units are displayed in the same sequence as the settings made from 81 S u to 85 S u 3 If 00 is set no unit of measurement will be...

Страница 20: ...ht unit Sample weight Total g g 1 500 500 pcs Measured number of samples If samples to be counted deviate widely in weight or a higher measure of accuracy is desired it is recommended that users use t...

Страница 21: ...e sample measurable unit weight is insufficient For the range of unit weights that can be measured and saved please refer to 13 Specifications on page 41 2 In the sampling of Operation Step 3 press th...

Страница 22: ...oximately an equal number of samples as the number currently displayed Add them gradually in groups of approximately the same number until the number of samples is approximately double The additional...

Страница 23: ...weight 1 Activating the percentage mode Press the Function key to disaplay 2 Setting the reference value Press the Set key The display begins flashing P SEt The balance is now ready for reference valu...

Страница 24: ...e saved limit weight please refer to 13 Specifications on page 41 2 While setting up the reference value in Step 2 the Set key has been pressed while the samples were loaded on the balance If L Err ap...

Страница 25: ...the limit value 2 Numeric value setup method The limit value is entered with a key stroke The limit values entered are held in memory even when the balance is powered down The respective limit values...

Страница 26: ...etting the judgement range Press the Function key The display changes to Judgement Range Press the Zero Tare key to select the desired choice 6 Setting the number of judgement points Press the Functio...

Страница 27: ...value has been saved the balance displays it briefly and proceeds to the following setup If One point setup was chosen the setup is complete 4 The upper limit value setup The display now changes to H...

Страница 28: ...moves to the digit on the immediate left Each time the key is pressed the flashing digit moves one position left When the leftmost digit is selected the flashing advances to the rightmost digit posit...

Страница 29: ...the Function key during the setup of actual quantities or the Set key during the setup of values 3 If you press the Memory key while a value is flashing an actual quantity will be set based on the wei...

Страница 30: ...plicable models HYPROP BALANCE Span adjustment should be performed with the balance installed perfectly level and without any load on the measurement pan 1 Activating the span adjustment function Pres...

Страница 31: ...cisely as possible use a weight close to the capacity of the balance Calibration weights can be purchased from Shinko Denshi or local dealers For detail information please contact Shinko Denshi or loc...

Страница 32: ...simultaneously Display will change to a flashing CAL 0 and the zero point will be calibrated automatically 3 Zero point calibration Once zero point calibration is completed display switches to CAL on...

Страница 33: ...r is turned on with the knob in the Calibration position or when set to the Calibration position during measurement 4 Wind or vibration affecting the balance may result in the process not proceeding b...

Страница 34: ...n of HYPROP BALANCE Series Page 29 GLP Compliant Print Form CALI BRATI ON MO DEL S N I D DATA TI ME CAL END NAME I D Title Model of balance Serial No of balance ID No Date of calibration Time of calib...

Страница 35: ...32 10 Input output functions 10 1 Terminal Numbers and Functions Discontinued...

Страница 36: ...33 10 2 Connection between Balances and Personal Computers Discontinued...

Страница 37: ...codes ASCII codes 8 bit 4 Signal levels Compliant with EIA RS 232C HIGH level Data logic 0 5 to 15 V LOW level Data logic 1 5 to 15 V 5 One character bit configuration Start bit 1 bit Data bit 8 bits...

Страница 38: ...D8 U1 U2 S1 S2 CR LF 3 Six digit numeric format for model provided with an auxiliary scale interval Composed of 15 characters including the terminators CR 0DH LF 0AH with added to the left of the auxi...

Страница 39: ...eger it may be omitted and replaced with a blank space SP in the lowest order place SP 20H Space zero of leading portion of value leading zero suppress 2FH Delimiter to be inserted to the left of the...

Страница 40: ...em commands can be transmitted regardless of their data transmission timing 2 When the balance has executed the received command it activates a normal end response or transmits the requested data via...

Страница 41: ...H Outputs once if stable and outputs continuously when unstable Even if the sample is not replaced output of the balance stops when it stabilizes after being output once 7 4FH 37H Pressing Memory key...

Страница 42: ...rent when the power is switched on Overcharging will also decrease battery life 2 When the balance is used for the first time after purchase the operating time may be shorter than when using a fully c...

Страница 43: ...tainer weight of sample A section of the mechanism is damaged Check the total weight Execute tare subtraction again Contact our Technical Service Division or your local dealer u Err is displayed A for...

Страница 44: ...its indication seven digits Segment height 16 5 mm With backlight 3 Calibration span adjustment Span adjustment with external weight 4 Overload indication o Err is displayed if weight capacity 9 inter...

Страница 45: ...200CE g 2200 0 1 0 01 ct 11000 1 0 1 oz 77 0 01 0 001 lb 4 8 0 001 0 0001 ozt 70 0 01 0 001 dwt 1400 0 1 0 01 Hong Kong 58 0 01 0 001 Singapore Malaysia 58 0 01 0 001 Taiwan 58 0 01 0 001 mom momme 58...

Страница 46: ...aiwan 37 5 187 5 1 32277 0 08267 1 20565 24 11306 1mom 3 75 18 75 0 13228 0 00827 0 12057 2 41131 1to 11 66380 58 31902 0 41143 0 02571 0 37500 7 5 unit tael Hong Kong tael Singapore Malaysia tael Tai...

Страница 47: ...44 UMS GmbH Gmunderstrasse 37 81379 Munich TEL 49 89 12 66 52 0 FAX 49 89 12 66 52 20 info ums muc de www ums muc de...

Отзывы: