EQUIPMENT CARE

Care of upholstery 

The upholstery material used on the top, leg rest, an 
 headrest is resistant to most medical stains.

For light cleaning:

1

  A solution of 10% liquid soap and clean water applied with a 

soft damp cloth will remove disinfection cleaner build-up.

2

  If necessary, a solution of liquid cleanser and water can be 

applied with a soft bristle brush. Wipe away residue with a 
water-dampened cloth.

For disinfection:

1

  Dampen a soft white cloth with a solution of standard bleach 

(sodium hypochlorite) or other chlorine-based cleaner*** 
and water; 10% bleach, 90% water.

2

  Rub gently.

3

  If necessary, allow the 1:10 diluted bleach (sodium 

hypochlorite) solution to puddle on the affected area or 
apply with a clean, soaked cloth for approximately 30 
minutes. Rinse with a water-dampened cloth to remove 
any remaining bleach concentration.

4

  Rinse with a water-dampened cloth to remove cleaner 

solution and allow thorough drying of material.

***See current CDC Guideline for Disinfection & Sterilization in 
Healthcare Facilities for bleach alternative cleaners.

In laboratory testing, upholstery 
protected with PreFixx

®

 finish 

was treated with the following 
disinfectants with little to no 
discoloration or damage to the 
upholstery. This testing may not 

reflect actual results in the field.*

Recommended Disinfectants

  Clorox

®

 Broad Spectrum Quaternary 

Disinfectant Cleaner

  Clorox

®

 Healthcare Bleach 

Germicidal Cleaner

 Clorox

®

 Healthcare Bleach 

Germicidal Wipes

  Clorox

®

 Healthcare EZ-KILL

®

 Wipes

WHEN THE USE OF STRONG CLEANING 
SOLUTIONS IS NECESSARY, TEST AN 
INCONSPICUOUS AREA TO ASSURE THAT 
DAMAGE TO UPHOLSTERY OR PAINTED 
SURFACES WILL NOT OCCUR.

CAUTION:

Proper Sterilization Instructions

Upholstery with PreFixx® 

Protective Finish

 Clorox

®

 Healthcare VersaSure

®

 

Cleaner Disinfection Wipes

 Clorox

®

 Hydrogen Peroxide 

Cleaner Disinfectant Spray

 Clorox

®

 Hydrogen Peroxide 

Cleaner Disinfectant Wipes

  Agar™ Powerquat

 Asepticare™

  Asepticare™ TB-II

  AVISTAT-D™ Ready To Use 

Spray Disinfectant Cleaner

  Biotrol BirexSE

®

  Bleach 1:5 (20% bleach)

  Bleach 1:9 (10% bleach)

  Bleach-Rite® Disinfecting Spray

 CaviCide™

 CaviCide1™

  CaviCide™ AF

Note: Immediately remove any fluid spilled on upholstery surface.

Antimicrobial: UMF Medical Upholstery providers outstanding 

protection in difficult medical and healthcare environments and 

contains an agent effective against bacterial and fungal microorganisms.

Care of painted surfaces 

A chemical acid-resistant paint is used, but extreme care must be 
taken not to use ammonia-based cleaners or discoloration of paint may 

occur. A damp cloth or mild liquid soap solution should be sufficient.

Care of bright metal surfaces 

All non-painted surfaces, chrome plated, or stainless steel should be 
wiped weekly with a clean damp cloth then buffed to a lustrous shine 
with a soft dry cloth.

Care of base 

The protective base is easily washable with mild liquid soap and water.  
A soft bristle brush may be used on scuffed stained areas.

  Diversey™ Accel

®

 

INTERVention

®

 Wipes

  Diversey™ Avert

®

 Sporicidal 

Disinfectant Cleaner

  ERC Performance Wipes

  McKesson Disposable 

Germicidal Surface Wipes

  McKesson Pro-Tech RTU 

Disinfectant Cleaner

 OPTIM

®

 1 Wipes

 OPTIM

®

 33TB

 Oxivir

®

 1 RTU

 Oxivir

®

 Five 16

 Oxivir

®

 TB

  OxyCide™ Daily 

Disinfectant Cleaner

  PDI Sani-Cloth

®

 AF3 

Germicidal Disposable Wipes

Содержание FUSIONONE 3001

Страница 1: ...TURE MAXIMUM PATIENT WEIGHT 500 LBS 226 KG UMFmedical com Ph 814 266 8726 Fax 814 266 1870 1316 Eisenhower Blvd Johnstown PA 15904 customerservice umfmedical com OWNER S MANUAL FUSIONONE POWER EXAM CHAIR 3001 3002 3003 3 2518 Rev A ...

Страница 2: ...CSA CAN CSA C22 2 NO 60601 1 14 IEC 60601 1 Edition 3 1 2012 The entire table is considered to be an applied part because the patient or practitioner could come in contact with any part of the table while in use It is noted that the only components designed to come in contact with the patient on a regular occurrence are the upholstered top leg pad and stirrups CLASSIFICATIONS APPLIED PARTS IMPORTA...

Страница 3: ...h as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the ME EQUIPMENT or ME SYSTEM including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result 6 There are no maintenance requirements specifically related to EMC all maintenance items are addressed in the important information section ENVIRON...

Страница 4: ...ne of the uniform field area in one orientation of the ME EQUIPMENT or ME SYSTEM b ME EQUIPMENT and ME SYSTEMS that intentionally receive RF electromagnetic energy for the purpose of their operation shall be tested at the frequency of reception Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS This test assesses the BASIC SAFETY and ESSENTIAL PERFOR...

Страница 5: ...ve a surge protection device in the primary power circuit may be tested only at 2 kV line s to earth and 1 kV line s to line s k Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS i Direct coupling shall be used m r m s before modulation is applied n The ISM industrial scientific and medical bands between 0 15 MHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 ...

Страница 6: ...pply Optional 220V 50 60HZ 5AMP Power Cord Length 10 FT 3 0m Table Weight 400 lbs 181kg Paper Roll Maximum size 21 0 Long x 3 5 Diameter 53 3 cm x 8 9 cm Storage Drawer Front 19 0 W x 12 0 D x 3 5 H 48 cm x 30 cm x 9 cm Storage Drawer Side 24 5 W x 12 5 D x 3 5 H 62 cm x 32 cm x 9 cm SPECIFICATIONS ...

Страница 7: ...A SEAT BACKREST IS NOT DESIGNED TO SUPPORT PATIENT S FULL WEIGHT WARNING Hand Foot Control Table Operation FEATURES AND OPERATION Hand Control Foot Control Table High Low Function Operation 1 Press the up button to raise the table 2 Press the down button to lower the table DO NOT PLACE FOOT REST ON BASE DURING OPERATION WARNING Optional accessory 581 ...

Страница 8: ... drawer for cleaning purposes 1 Slide leg section forward until it contacts the two stopper pins 2 When not in use push the leg section in to stow 3 Leg rest capacity should not exceed 100lbs DO NOT USE DRAWER OR DRIP PAN AS A SEAT OR STEP WARNING DO NOT USE THE LEG REST AS A SEAT OR STEP WARNING FEATURES AND OPERATION Products placed in warming drawer should be checked for proper temperature befo...

Страница 9: ...Release stirrup to lock into position 1 Lift up the seat and the pelvic tilt rod will automatically fall into place 2 To disengage pelvic tilt lift seat and flip the lever on either side of the seat back ENSURE PELVIC SUPPORT ARM IS LOCKED AND HANDS ARE NOT UNDERNEATH BUMPER WHEN LOWERING SEAT SECTION DO NOT SUPPORT THE PATIENT S ENTIRE WEIGHT WITH THE STIRRUPS ENSURE STIRRUPS ARE PROPERLY LOCKED ...

Страница 10: ...e panels Step 3 Slide cages to opposite side of table Step 5 Install side panel on opposite side Step 2 Remove drawer Extend drawer Press tabs on both sides Remove drawer Step 4 Align slides and install drawer on opposite side Slide in until an audible click is heard ...

Страница 11: ... During normal operation the front drawer and its contents should be between 100 F and 110 F If temperatures exceed 120 F contact UMF Medical Customer Service COMMON OPTIONS AND ACCESSORIES THE HEATING ELEMENT FOR THE DRAWER WARMER IS LOCATED IN A PANEL ABOVE THE DRAWER THIS PANEL WILL BE HOT WHILE DRAWER WARMER IS ON AND IS NOT TO BE TOUCHED WARNING ...

Страница 12: ...ULD NOT BE USED FOR ANY OTHER ITEMS OUTLET IS FOR MEDICAL EQUIPMENT ONLY CAUTION CHECK GROUND CONTINUITY PERIODICALLY 120V MAXIMUM OUTPUT 5 0A 120V 220V MAXIMUM OUTPUT 2 5A 220V WARNING DO NOT DISCARD COVER COVER IS TO BE REAPPLIED AFTER REMOVAL OF EQUIPMENT FROM MULTIPLE SOCKET OUTLET WARNING WARNING GROUND RELIABILITY CAN ONLY BE ACHIEVED WHEN POWER SUPPLY IS CONNECTED TO AN EQUIVALENT RECEPTACL...

Страница 13: ...ES Bierhoff Knee Crutch 251 Pair 1 Extend stirrups to full extended position with heel stirrup in retracted position 2 Insert knurled end of knee crutch rod into hole on end of stirrup 3 Adjust to position and tighten with slide lock ...

Страница 14: ...ABLE CONNECTION Plug Power Cable into Column Flip Cable Lock Down Plug Control Cable Into column Note Make sure to position table such that the access to the cable connection is not impeded under normal use ...

Страница 15: ... s power supply When plugged in the indicator light should be illuminated indicating the power supply is operating correctly Note The table does not have an on off switch so whenever the power supply is plugged in and connected to the column the table is operable ...

Страница 16: ...S NECESSARY TEST AN INCONSPICUOUS AREA TO ASSURE THAT DAMAGE TO UPHOLSTERY OR PAINTED SURFACES WILL NOT OCCUR CAUTION Proper Sterilization Instructions Upholstery with PreFixx Protective Finish Clorox Healthcare VersaSure Cleaner Disinfection Wipes Clorox Hydrogen Peroxide Cleaner Disinfectant Spray Clorox Hydrogen Peroxide Cleaner Disinfectant Wipes Agar Powerquat Asepticare Asepticare TB II AVIS...

Страница 17: ...fessional No Rinse Sanitizing Multi Surface Wipes SaniZide Plus STERI 7 XTRA CONCENTRATE recommended dilution ratio 1 10 STERI 7 XTRA WIPES Vert 2 Go ED Virox AHP 5 Virox PREempt RTU Virex II 256 Virex Plus One Step Disinfectant Cleaner Deodorant Wayne Concept 256N Wex Cide 128 Care Cleaning Use one of the following cleaners with a soft cloth or damp sponge Rinse area with fresh water then dry wit...

Страница 18: ...ns caused in whole or in part by mishandling misuse abuse negligence alteration accident freight damage tampering or failure to seek and obtain repair or replacement in a timely manner Products which are not installed used and properly cleaned and maintained as required in the UMF Medical installation and or Owner s Manual for the applicable product Replacement parts alterations or installation of...

Страница 19: ...19 This page intentionally left blank ...

Страница 20: ...UMFmedical com Ph 814 266 8726 Fax 814 266 1870 1316 Eisenhower Blvd Johnstown PA 15904 customerservice umfmedical com ...

Отзывы: