SAFETY CAUTIONS
iii
Operating environment
Do not connect the sensor head to a test object of which pressure is in excess of
atmospheric pressure. If the pressure in the sensor head exceeds atmospheric
pressure, the sensor head will be damaged or it will pop out from the connector,
causing injury to the surrounding, including human body. If the pressure exceeds
atmospheric pressure, provide an isolation valve so that the pressure in the sensor
head does not exceed atmospheric pressure.
Operating environment
Do not use the gauge in a place where it may be splashed with water. If it is
splashed with water, failure, earth leakage or fire can result.
Ventilation
Do not plug the air vents of the vacuum gauge controller. If the air vents are
plugged, heat will be contained inside and the gauge may be damaged.
Keep out foreign matter.
If foreign matter like metal fragments or combustibles are admitted into the
vacuum gauge through the air vents or other openings, remove them. Otherwise,
the vacuum gauge may be damaged.
Operating conditions
Operate the vacuum gauge under the environment set forth in the specifications.
Repacking for transfer
If the vacuum gauge is to be shipped to other site, repack it in the same way as on
delivery. If the gauge is shipped bare, it may be damaged.
Disposal
When discarding the vacuum gauge, comply with your local regulations. If the
gauge was used in an environment that can cause hazards to human body, have
it disposed of by an authorized specialist in disposal.
Customer shall be responsible for the cost relating to disinstallation.
Meter panel
Don’t long contact with chemical cleaning cloth, don’t wipe with benzene or
thinner. The pointer may become unstable due to static electricity. In the case,
apply a commercially available antistatic agent to the cover.
Connection
The contact output capacity is 100VAC/0.5A, 24VDC/1A (resistance load).
If power higher than this is opened/closed, do not use the contact of the vacuum
gauge, but use a large capacity switch in conjunction.
✔
Note
Maintenance
Aluminum electrolytic capacitor is used for the electric circuit in the sensor unit.
Generally, the life expectancy of the aluminum electrolytic capacity is limited
and the higher the surrounding temperature, the shorter the life. It is
recommended to replace the aluminum electrolytic capacitor once every five
years or at the time or repair or overhaul to prevent components from being
damaged.
✔
Note
Lifetime of sensor head cable and connector
Female connector is used in the sensor head cable and detection unit. This
female connector has a lifetime. A large current load may cause a fire in the
connector connection area. Avoid the frequent insert/remove of connector to
prevent a heavy load and loosens of the connector connection.
Содержание GP-1GRY
Страница 2: ......
Страница 22: ...OPERATING PROCEDURE 14 Fig 6 Pressure vs recorder output curve...
Страница 33: ...CERTIFICATE OF DECONTAMINATION 25 7 CERTIFICATE OF DECONTAMINATION...
Страница 36: ...RELATED DRAWINGS 28 Fig 11 Dimensional drawing for controller...
Страница 37: ...RELATED DRAWINGS 29 Panel cut dimensional Fig 12 Dimensional drawing for A type meter...
Страница 38: ...RELATED DRAWINGS 30 Panel cut dimensional Use until 0 8 Fig 13 Dimensional drawing for B type meter...
Страница 39: ...RELATED DRAWINGS 31 Panel cut dimensional Fig 14 Dimensional drawing for C type meter...
Страница 40: ...RELATED DRAWINGS 32 Fig 15 Dimensional drawing for measuring unit GP H...
Страница 41: ...RELATED DRAWINGS 33 Fig 16 Dimensional drawing for measuring unit GP BH...
Страница 44: ...RELATED DRAWINGS 36 Fig 20 Sensor head filament connection diagram...