- 16 -
4.2.
BCD Output
Digital signals of pressure values are outputted from the I/O connector on the rear panel of the
controller.
Refer to Table. 4-1 "I/O Connectors".
The relationship between mantissa parts A, B and C is as shown below.
SYMBOL
.
× 10
EXPONENT PART C
MANTISSA PART B
MANTISSA PART A
The above are outputs during measurement.
The following are outputs at other displays.
(1)
When 0.0
×
10
-1
Pa is displayed
: 0.0
×
10
-1
When 0.0
×
10
-3
Torr is displayed
:
0.0
×
10
-3
(2)
When 2.7
×
10
3
Pa is displayed by blinking
: F.F
×
10
+F
When 2.0
×
10
+1
Torr is displayed by blinking
: F.F
×
10
+F
(3)
When E.E
×
10
E
is displayed by blinking
: E.E
×
10
+E