SK00-9662-EQ-003-00 

 

 

G-TRAN SERIES 

Pirani Vacuum gauge SWU10-R 

Quick Manual

 

 

Introduction 
This quick manual is for quick check of operation and display of the product. 
Please  refer  to  instruction  manual  in  advance  for  detailed  information  about 
operation, precautions and safety for proper use.

 

Available for download from 

ULVAC website.

 

https://showcase.ulvac.co.jp/ja

 

 

This manual is for the following gauges. Serial Nos. 00001 and higher

.

 

 

 

Part Names and Functions

 

SWU10-R controller section 

 

 

 

②           

   

  ③

 

Name 

Function 

 

I/O connector 

The I/O connector for the power supply and signals 
(RJ-45 jack connector). 

 

Gauge head 
clamp 

The clamp to hold and lock the gauge head. 

 

Mating screws 

Mating screws (M3) for the clamp hood. 

 

SWU10-R I/O connector 

 

No 

This unit 

Connection destination 

Power 5 VDC 

SH2/ST2 

Multi-ionization gauge 

GND 

Power 5 VDC 

Communication terminal 

Communication terminal 

GND 

Power 5 VDC 

GND 

 

SWU10-R LED 

 

Name (symbol) 

Function 

 

POWER/ERROR LED 

On        : Operating normally 

Flashing: Error such as a filament 
break, this unit is malfunctioning 
Off        : Power off, this unit or the 
SH2 / ST2 is malfunctioning 

 

 

Attaching the Product 

Loosen the clamp locking screws to loosen the clamp. Then attach the gauge 

head with the procedure below. 

 

Align the pins for the gauge head and push it all the way in 

 

Clamp the gauge head with the clamp 

 

Tighten the locking screws to secure the gauge head 

 

 

 

Handling Precautions 

3.1.

 

Precautions related to usage 

・ 

Secure  cables  as  much  as  possible  so  that  force  is  not  applied  to  cable 
connectors. 

・ 

Firmly insert the RJ-45 connector. 

・ 

To perform accurate pressure measurements, allow the pressure sensor to 
age for 20 minutes or longer before measuring. 

 
 

3.2.

 

Attaching this unit 

・ 

Attach  this  unit  so  that  the  gauge  head  attachment  opening  surface  is 
parallel to the gas flow. In particular, ensure that gases do not enter the 
gauge head interior like a beam. 

・ 

As much as possible, attach the gauge head so the filament is vertical to 
the force of gravity.   

・ 

The Pirani vacuum gauge head filament is thin at φ25 µm, so avoid use in 

a  location  with  a  large  amount  of  vibration  as  much  as  possible.  The 
biggest cause of filament breaks is from mechanical shock, so use caution 
regarding the installation location and handling. 

・ 

Use an O-ring to attach the gauge head that releases little gas. There is a 
risk  of  measurement  errors  or  the  gauge  head  lifespan  will  decrease  if 
materials that release a large quantity of gas, such as rubber tubes or grease, 
are used in the gauge head connection. 

 

 

 

Gas flow direction 

 

 

Checking the Filament and Temperature Sensor for 

Burnout 

Filament 
Approx. 5 

Ω

 

Temperature 
sensor  
Approx. 1.1 
k

Ω

 

The diagram below shows the connection of the Pirani sensor head filament. Check 

continuity referring to this diagram. The resistance of the filament is about 5Ω under 

atmospheric pressure and that of the temperature se

nsor is about 1.1 kΩunder 

atmospheric pressure. 
Note)Beware of over-current 

If the filament or temperature sensor is energized with an over-current of 2 mA or 

more, it may burn out. Be alert to the current value on the measuring instrument. 
Also measure the resistance value under atmospheric pressure because in vacuum, the 
filament and other components are heated and an accurate resistance value cannot be 
measured. 
 

 

Specifications

 

Name 

Pirani vacuum gauge measuring unit 

Model 

SWU10-R 

Measurement principle 

Pirani vacuum gauge that uses thermal 
conduction by gas molecule 

Number of gauge heads 

1pc 

Applicable gauge head 

SWP-16,SWP-R1/8,SWP-P18,SWP-P15,  SWP-
25, SWP-CF16, SWP-1S 

Measurement pressure range 

5.0×10

-2

 to 1.0×10

+5

Pa 

Output pressure range 

5.0×10

-2

 to 1.0×10

+5

Pa 

Accuracy (N

2

1×10

-1

Pa to 1×10

+4

Pa: ±10% 

5×10

-2

Pa to 1×10

-1

Pa: ±20% 

1×10

+4

Pa to 1×10

+5

Pa: ±20% 

Repeatability 

1×10

-1

Pa to 1×10

+4

Pa: ± 2% 

Sampling time 

50 ms, 5x moving average 

Measurement value output 

Output by serial communication 

LED display 

POWER/ERROR LED 

Gauge head material 

Filament

Pt 

Other

SUS304, FeNiCo, Ni, Au, Glass, Ceramic 

Gauge head withstand 
pressure 

2×10

+5

Pa(abs) 

*Take the withstand pressure for flanges, 
clamps, and other components into account 
separately. 

Operating temperature range 

10 to 40

 

Operating humidity range 

15 to 85% RH (no condensation) 

Storage temperature 

-20 to 65

(no condensation) 

Bake out 

150

(without electronics unit) 

Mounting position 

Free 

IP protection class 

IP30 

Power supply voltage 

5VDC

±

5% 

I/O connector 

RJ-45 jack connector 

Gauge head internal volume 

Pirani vacuum gauge head(SWP-16): 7.3cm

3

 

Weight 

Controller: Approx. 38 g, 

External dimensions 

φ

40

×

51 (controller section only) 

 

Standard Accessories 

Quick manual(This paper) 

1copy 

 

 

Trouble shooting 

Bad electrical wiring is a major cause of this unit malfunctions. If this unit 

behaves in a manner you find unusual, first check the wiring between this unit and 
the SH2/ST2. Refer to the SH2/ST2 instruction manual as well. 

 

The POWER/ERROR LED is off even though power is being supplied 

Cause 

Action 

Miswired or broken cable 

Repair the wiring and check continuity 
with a tester. 

The power supply voltage to the 
SH2/ST2 is not within the specified 
range 

Check the power supply voltage with a 
tester. 

This unit or the SH2/ST2 is 
malfunctioning 

Requires inspection and repairs at 
ULVAC. 

 

 

The POWER/ERROR LED is flashing or the SH2/ST2 keeps outputting the 
error signal 

Cause 

Action 

Filament break or short circuit 

Measure filament continuity and 
isolation between each pin and the case. 

Replace the gauge head when there is 
a problem. 

The gauge head is not attached 

Attach the gauge head. 

This unit or the SH2/ST2 is 
malfunctioning 

Requires inspection and repairs at 
ULVAC. 

 

 

Pressure does not reach 5

×

10

-2

 Pa or lower even when pumping to vacuum 

Cause 

Action 

The actual pressure is not 5

×

10

-2

 Pa 

or lower 

Check the pressure with an ionization 
vacuum gauge. 

Room temperature, gas temperature 
is different when measuring 

This is normal. Measurements are 
temperature dependent. 

This unit or the SH2/ST2 is 
malfunctioning 

Requires inspection and repairs at 
ULVAC. 

 

 

The pressure reading does not show a steady value. 

Cause 

Action 

The pressure is actually fluctuating 

This is normal. 

This unit is affected by pump 
vibrations and fluctuating 

Suppress vibrations as much as possible. 
Measurements are affected by the 
principle of the Pirani vacuum gauge. 

Attached to a location with a gas 
flow 

Attach this unit to a location with no gas 
flow. 
Measurements are affected by the 
principle of the Pirani vacuum gauge. 

The environmental temperature is 
fluctuating 

Wait for the environmental temperature 
to stabilize. 
Measurements are affected by the 
principle of the Pirani vacuum gauge. 

High temperature and low 
temperature gases are flowing in 

Implement corrective action such as 
attaching the gauge head to a location 
unaffected by gas temperature. 
Measurements are affected by the 
principle of the Pirani vacuum gauge. 

Dirty gauge head interior 

Replace the gauge head 

Affected by moisture and dirt 

Attach the gauge head to a location as 
unaffected as possible 
Influenced by heat of vaporization by 
gradual vaporization. 

There is a leak in the gauge head 
or the area where the gauge head is 
attached 

Gauge head leak 

  Replace the gauge 

head 
Other leaks 

  Stop the leak 

Bad cable contact, increased wire 
resistance 

Check the cable connection and 
resistance. 

The cable is affected by 
electromagnetic induction. 
(By external noise) 

Change the cable installation location. 
Use this unit with the device generating 
the noise turned off. 
Review how the cables are installed 
Examine separate measures for noise 

 

 

1

8