Chapter 3: System Installation
00.053.204, Revision A
SonixTouch Q+ Service Manual
3-4
3.1.5
Wiring Requirements
3.1.5.1 Main AC Connection
The electrical feed to the system should be a dedicated/isolated line (no other equipment on the same line) with a
third-wire ground. Ensure a low impedance path for current to return to the source.
3.1.5.2 Ethernet (Hard-Wired) Network Connection
It is the user’s responsibility to provide an Ethernet connection to the system and to install the correct type of cable
in accordance with the building's applicable standards. The commonly used cable is CAT5 (Category 5, 10 Base-T,
unshielded twisted pair).
3.1.5.3 Image Management Network
Obtain the following information from the system administrator:
• list of all equipment that is part of the Image Management network
• logical diagram of the network showing topology, subnets, etc.
• location of all equipment
• location of all LAN (Local Area Network) attachment points
• locations of all power outlets and connector types
• location of any dedicated analog phone line
• all necessary Internet Protocol (IP) addresses and subnet information.
Caution:
System networking options are intended for use inside your organization's firewall. Organizations that
elect to configure/use the networking functionality provided by Ultrasonix are assuming all liabilities and risks
associated with that decision.
Содержание SonixTouch Q+
Страница 1: ...Analogic Ultrasound Service Manual SonixTouch Q Ultrasound System...
Страница 2: ......
Страница 4: ......
Страница 14: ...Chapter 2 System Specifications 00 053 204 Revision A SonixTouch Q Service Manual 2 2...
Страница 42: ...Chapter 4 Performance Testing 00 053 204 Revision A SonixTouch Q Service Manual 4 4...
Страница 120: ...Chapter 8 DICOM 00 053 204 Revision A SonixTouch Q Service Manual 8 24...
Страница 147: ......