20
NT UF 831 CO GB (b) / March 2014
1) Analog inputs/outputs
The UF 831 allows two types of inputs/outputs to be used:
Current inputs/outputs (0-20mA, 4-20mA, 0-24mA) or voltage inputs/outputs (0-10V),
Temperature measurement Pt 100 and Pt 1000 with 2, 3 or 4 wires.
The sensors must be powered externally. The wiring of the analog inputs/outputs is indicated in appendix 5:
Wiring of the inputs - outputs.
2) On/Off inputs/outputs (relay)
Each relay of the UF 831 is a normally-open relay contact:
Breaking capacity of each relay: U < 50 V, I < 10 mA,
The maximum frequency of a relay in pulse mode depends on the length of the pulse.
3) Serial link
The serial link is connected using a connector, which can be configured by RS232 or RS485 wiring. The
wiring diagrams are presented in appendix 5: Wiring of the communication port.
4) Connection to the PC
The connection to the PC is made using a USB connector on the front panel of the flow meter.
Содержание UF 831
Страница 5: ...5 NT UF 831 CO GB b March 2014 CHAPTER 1 OVERVIEW...
Страница 10: ...10 NT UF 831 CO GB b March 2014 CHAPTER 2 SAFETY INSTRUCTIONS...
Страница 14: ...14 NT UF 831 CO GB b March 2014 CHAPTER 3 INSTALLATION AND WIRING...
Страница 22: ...22 NT UF 831 CO GB b March 2014 CHAPTER 4 IMPLEMENTING A MEASURING POINT...
Страница 26: ...26 NT UF 831 CO GB b March 2014 CHAPTER 5 USING AND CONFIGURING THE UF 831...
Страница 59: ...59 NT UF 831 CO GB b March 2014 CHAPTER 6 FUNCTION ENGINE...
Страница 61: ...61 NT UF 831 CO GB b March 2014 CHAPTER 7 PC SOFTWARE...
Страница 77: ...77 NT UF 831 CO GB b March 2014 Appendix 1 Characteristic echo signals...
Страница 80: ...80 NT UF 831 CO GB b March 2014 Appendix 2 Speed of sound in water...