Instruction for Outer Sheath Cleaning and Sterilizing 

•  The cleaning should be performed no later than 24 hours after treatment. The outer sheath can be cleaned by 

automated cleaning.

•  The outer sheath is not sterile upon receipt and must be steam autoclave sterilized prior to use in accordance with 

the following instructions.

1.  After removing the outer sheath from the motor component, carefully place the outer sheath onto the disinfection 

tray. To prevent scratching of the surface and cross contamination of sheath, do not allow products to touch each 

other, or place into a sterilization bag.

2.A 

For manual cleaning

•  Use tap water, purified, distilled, or deionized water at room temperature of 25°C to pre-rinse for 3 minutes to 

remove any visible surface contaminants.

•  Clean with cleaning agent at the condition recommended by the cleaning agent manufacturer for 5 minutes: 

Detergent: RUHOF ENDOZIME® AW PLUS WITH APA Dilution Ration: 1:270 Temperature: <60°C Contact Time: 5 

minutes.

2.B 

For automated cleaning

 

•  Use only washer-disinfectant that meets the FDA requirements for cleaning processes and follow the 

manufacturer’s instructions for correct use.

•  Wear appropriate PPE (e.g., mask, protective eyewear, and gown) when splashing or spraying is anticipated 

during cleaning.

•  Use only verified cleaning agents below to clean the outer sheath:

•  Cleaning agent name: RUHOF ENDOZIME

®

 AW PLUS WITH APA

•  Manufacturer: The RUHOF CORPORATION

3.  Rinse away any contaminants with purified water, distilled, or deionized water at room temperature of 25°C 

for 1 minute.

4. Use a clean sterile cloth or dry paper to wipe the surface to remove excess water.  

5.  Package the cleaned/disinfected and dried outer sheath in a sterilization pouch to maintain sterility. It is 

recommended to use a legally marketed pouch or other method of maintaining sterility for the following 

sterilization cycle.

Note:

•  Avoid contact with different metals when packaging.

•  Cold liquid disinfection/sterilization, chemical vapor sterilization, and dry heat sterilization methods have not 

been tested or validated for efficacy and are not recommended for use.

•  Do not immerse the outer sheath in an ultrasonic bath.

•  The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5) will reduce the life span of products. 

Note:

 It may be advised to bag sheath in a sterilization pouch prior to sterilizing to reduce discoloration and 

Содержание Ultrapro TX Cordless

Страница 1: ...Instructions for Use...

Страница 2: ...Contents...

Страница 3: ...for Use Intended Purpose 5 4 Contraindications 6 5 Warnings and Precautions 6 6 Stepwise Instructions 7 7 Maintenance 9 8 Processing 10 9 Storage and Disposal 15 10 Technical Considerations 15 11 Misc...

Страница 4: ...4 Disposable prophy angle Centralized control button Outer sheath Motor component Battery indicator Disposable sleeve Power adapter Foot pedal Charging station Charging indicator...

Страница 5: ...dental prophylaxis procedures and is compatible with nearly all disposable prophy angles on the market Product Components 1 Motor component 3 Autoclavable outer sheaths 1 Charging station 2 Power ada...

Страница 6: ...ces handheld or mobile high frequency communication devices Long term use of equipment may result in overheating of the motor handpiece and should be left to cool until next use Please contact Ultrade...

Страница 7: ...ht on the charging station and foot pedal will flash green Once fully charged the battery indicator light will illuminate green Note Handpiece and foot pedal are provided partially charged charge leve...

Страница 8: ...shut off automatically charge immediately Bluetooth Indicator Charging Socket Battery Indicator BATTERY INDICATOR S COLOR BATTERY OF FOOT PEDAL STATUS Green light flashes Charging Green light illumin...

Страница 9: ...tal channel Pull the old O ring off and replace with a new O ring Warranty Ultradent hereby warrants that this instrument shall for a period of 2 years conform in all material respects to the specific...

Страница 10: ...instruments where applicable after sterility PART NAME INFECTION PREVENTION Disposable Prophy Angle DPA Prophy angle that helps clinician easily access hard to reach areas of the mouth quickly for opt...

Страница 11: ...the motor component carefully place the outer sheath onto the disinfection tray To prevent scratching of the surface and cross contamination of sheath do not allow products to touch each other or plac...

Страница 12: ...ses and follow the manufacturer s instructions for correct use Wear appropriate PPE e g mask protective eyewear and gown when splashing or spraying is anticipated during cleaning Use only verified cle...

Страница 13: ...sterilization time of 4 minutes at a temperature of 132 134 and a pressure of 185kPa to 190kPa with a drying time of 20 minutes 3 The highest sterilization temperature is 138 4 Allow a maximum sterili...

Страница 14: ...station etc until the surface is clean 2 Wipe the surface of the component with a dry soft nap free cloth 3 Repeat the above steps at least 3 times or until all visible soil is removed 4 Wipe surfaces...

Страница 15: ...cal waste according to local laws and regulations The motor component and foot pedal include a battery pack lithium ion battery Do not dispose of in a fire or source of heat Doing so may cause rupture...

Страница 16: ...inutes Troubleshooting FAILURE SOLUTION Disposable prophy angle is not rotating 1 Ensure that the outer sheath and the disposable prophy angle are attached securely 2 Ensure that the motor component a...

Страница 17: ...in the hole of the foot pedal and the pressed handpiece button to ensure a Bluetooth connection In the case that this does not work on the first attempt repeat the steps until there is a Bluetooth co...

Страница 18: ...omagnetic environment guidance RF emissions CISPR 11 Group 1 The Ultrapro Tx Cordless handpiece uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely...

Страница 19: ...1000 4 5 0 5 1kV line to line 0 5 1 2kV line to earth 0 5 1kV line to line 0 5 1 2kV line to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short...

Страница 20: ...ended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b In...

Страница 21: ...d 1 2 P1 2 800MHz to 2 7GHz d 2 3 P1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommende...

Страница 22: ...mitation indicates the maximum and minimum temperature limits at which the item shall be stored and transported Caution Federal law restricts this device to sale by or on the order of a licensed denta...

Страница 23: ...23...

Страница 24: ...Manufactured for Ultradent Products Inc Made in China 505 West Ultradent Drive 10200 South South Jordan Utah 84095 1011506AR01 092222...

Отзывы: